these should be removed. Use the new valve as a reference to determine which two screws need to be removed.

8.Note the orientation of the valve body on the manifold so that the new valve can be installed in the same orientation.

9.Remove valve body from sensor manifold.

10.Clean any residue or dirt off the surface of the manifold using a lint-free cloth and distilled or DI water.

11.Install the new valve by reversing steps 1-7.

9.8.Configuring the Analog Output

The analog output can be configured for either 0-5V DC or 4-20mA operation. To change or verify the configuration of the analog output:

1.Disconnect power from the M465M.

2.Remove the six screws and the top cover (Rack Mount Configuration) or open front panel (NEMA Configuration.)

3.Locate the Mainboard PCA (see Figure 9-1or Figure 9-2.)

4.Set the desired operation as shown in Figure 9-4.

5.Re-Install the top cover or re-secure the front panel (NEMA Configuration.)

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 58

Image 60
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsSafety ApprovalsEMC Other Type CertificationsPrinciple of Operation 3  Page Mechanical Installation for Nema 4X Enclosure Getting StartedVentilation Clearance Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Location of I/O Connectors in the Nema Configuration Electrical I/O ConnectionsAnalog Output Status Outputs Status Name Normal Operating Output # StateHi-Current Relay Outputs Digital Serial InterfacesRelay # Function Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Low Flow Alarm Optional Hardware and SoftwareRack Mount Slides This page is Intentionally Left Blank Operation Front Panel OverviewOzone Concentration Display Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStatus LED’s Status LED Normal StateFront Panel Menus Menu Navigation BasicsEditing Numeric Values Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeSETUP-VARS Menu Meaning Default ValueRange SETUP-ALARM Menu Config MenuProperty Meaning Default Range Value This page is Intentionally Left Blank Changing COM port Hardware Configuration Digital CommunicationsDigital Communications Overview RS232 Configuration RS-232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCEProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEEthernet Configuration Hardware Setup for Configuring the Ethernet ModuleVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Instrument Setup and Configuration Registers Read / Write Modbus Register MapsReal-Time Concentration and Status Registers Read Only Decimal Star Address Description Register Type Bit BitsMaintenance Maintenance ScheduleMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement UV Lamp Adjustment UV Lamp Replacement1150 mV Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Operation Boot Loader Hardware SetupBoot Loader Software Setup Upload Firmware Page Troubleshooting Pneumatic DiagramReference Drawings Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Pneumatic Error Sensor OKInvalid Reading Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version