Figure 4-4: Pneumatic Connections, NEMA Configuration

4.4.1.Ozone Inlet Connection

Sample gas connections are made to the ¼” stainless steel compression fitting on the rear panel. (See Figure 4-3and Figure 4-4)

1/4" O.D. FEP (Teflon) or Stainless Steel tubing is recommended to connect the sample source to the monitor. Any fittings used in the sample lines should be constructed of Stainless Steel or Teflon.

The M465M has an internal sample pump and is designed to sample gas that is close to ambient pressure. Ensure that the sample gas pressure is in the range of 11 – 16 psia (pounds per square inch absolute.)

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 20

Image 22
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsEMC SafetyApprovals Other Type CertificationsPrinciple of Operation 3  Page Ventilation Clearance Mechanical Installation for Nema 4X EnclosureGetting Started Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Electrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status Outputs Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsDigital Serial Interfaces Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Optional Hardware and Software Low Flow AlarmRack Mount Slides This page is Intentionally Left Blank Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadStatus LED’s Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Status LED Normal StateMenu Navigation Basics Front Panel MenusEditing Numeric Values Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeMeaning Default Value SETUP-VARS MenuRange Config Menu SETUP-ALARM MenuProperty Meaning Default Range Value This page is Intentionally Left Blank Digital Communications Changing COM port Hardware ConfigurationDigital Communications Overview RS232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-232 ConfigurationProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Real-Time Concentration and Status Registers Read Only Instrument Setup and Configuration Registers Read / WriteModbus Register Maps Decimal Star Address Description Register Type Bit BitsMaintenance Schedule MaintenanceMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement UV Lamp Replacement UV Lamp Adjustment1150 mV Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Hardware Setup Boot Loader OperationBoot Loader Software Setup Upload Firmware Page Pneumatic Diagram TroubleshootingReference Drawings Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version