10.2.4.1. Pressure Too High

The monitor inlet is being pressurized. The monitor inlet should be allowed to sample gas at ambient pressure. Disconnect all sample lines from the monitor and read the pressure again. If the value then drops to within the acceptable range, then one of the sample lines is at elevated pressure.

If the inlet is being pressurized, then the design of the sample system should be reviewed and corrected.

10.2.4.2. Pressure Too Low

Something may be restricting the flow through the monitor, causing the sample pump to pull a vacuum on the sensor module. First, unplug the sample pump power plug on the Mainboard PCA and observe the pressure reading. If the pressure reading is still too low, then likely there is a problem with the pressure sensor on the Sensor Module and the Sensor Module should be replaced. Next reconnect the pump and disconnect all sample lines from the monitor and see if the pressure increases to normal levels. If it does, then the problem is somewhere in the sampling system external to the monitor. If the pressure is still below the limit with the sample lines removed, then there is a restriction in the pneumatics of the monitor upstream of the sensor assembly. The most likely cause is a plugged particulate filter. Disconnect the outlet fitting on the sample filter and observe the pressure. If it returns to normal levels then the restriction is in the particulate filter or inlet manifold assembly (multi-stream configurations only.) Replace particulate filter per Section 9.3.

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 68

Image 70
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsEMC SafetyApprovals Other Type CertificationsPrinciple of Operation 3  Page Ventilation Clearance Mechanical Installation for Nema 4X EnclosureGetting Started Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Electrical I/O Connections Location of I/O Connectors in the Nema ConfigurationAnalog Output Status Outputs Status Name Normal Operating Output # StateRelay # Function Hi-Current Relay OutputsDigital Serial Interfaces Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Optional Hardware and Software Low Flow AlarmRack Mount Slides This page is Intentionally Left Blank Ozone Concentration Display OperationFront Panel Overview Menu Navigation KeypadStatus LED’s Concentration Alarm Status/ResetAlarm Status LED Color Alarm State Status LED Normal StateMenu Navigation Basics Front Panel MenusEditing Numeric Values Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeMeaning Default Value SETUP-VARS MenuRange Config Menu SETUP-ALARM MenuProperty Meaning Default Range Value This page is Intentionally Left Blank Digital Communications Changing COM port Hardware ConfigurationDigital Communications Overview RS232 Configuration Function DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-232 ConfigurationProperty Value RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEVerifying Network Hardware Connection Ethernet ConfigurationHardware Setup for Configuring the Ethernet Module Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Real-Time Concentration and Status Registers Read Only Instrument Setup and Configuration Registers Read / WriteModbus Register Maps Decimal Star Address Description Register Type Bit BitsMaintenance Schedule MaintenanceMaintenance Item Recommended Interval Section Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement UV Lamp Replacement UV Lamp Adjustment1150 mV Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Hardware Setup Boot Loader OperationBoot Loader Software Setup Upload Firmware Page Pneumatic Diagram TroubleshootingReference Drawings Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Invalid Reading Pneumatic ErrorSensor OK Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version