10.2.Troubleshooting Using Front Panel Status LED’s or Status Outputs

The M465M has 4 front-panel status LED’s that reflect the current operating status of the monitor, and indicate fault conditions. There are also four relay Status Outputs on the rear panel (or internally in the NEMA configuration) that also reflect the state of these status LED’s. See Section 4.5.3 for more information on the Status Outputs.

Table 10-1: Status LED/Output Definitions

Status

Status LED

Normal

Triggers

Critical

Output #

Label

Operating

 

 

Warning?

 

 

State

 

 

 

1

Sensor OK

On

Reference < 125.0 mV

Yes

 

 

 

Reference > 1230.0 mV – 0.5 Wt%

 

2

Invalid Reading

Off

O3 Concentration < F.S. Range

No

 

 

 

O3 Concentration > F.S. Range

 

3

Check Lamp

Off

Reference < 250.0 mV

No

4

Pneumatic

Off

Pressure < 9.0 psia

No

 

Error

 

Pressure > 18.0 psia

 

 

 

 

Flow < 500 cc/min*

 

*If Flow

Switch installed

 

 

 

 

P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 66

Image 68
Contents Model 465M Ozone Monitor This page is Intentionally Left Blank Safety Messages General Safety HazardConsignes DE Sécurité Mise EN GardeWarranty This page is Intentionally Left Blank Table of Contents Calibration Figures Tables Introduction PrefaceThis page is Intentionally Left Blank Specifications and Agency Approvals SpecificationsSafety ApprovalsEMC Other Type CertificationsPrinciple of Operation 3  Page Mechanical Installation for Nema 4X Enclosure Getting StartedVentilation Clearance Area Minimum required clearanceMechanical Installation 19 Rack Version M465M Mounting Hole Locations and DimensionsAC Power Connection for Monitors in Nema 4X Enclosures Nema AC Power Connection to Monitor Pneumatic Connections, Rack Mount Configuration Pneumatic ConnectionsPneumatic Connections, Nema Configuration Ozone Inlet ConnectionExhaust Connection Rear Panel Electrical I/O ConnectionsNema Electrical I/O Connections Analog Output Location of I/O Connectors in the Nema ConfigurationElectrical I/O Connections Status Outputs Status Name Normal Operating Output # StateHi-Current Relay Outputs Digital Serial InterfacesRelay # Function Normal Operating StateEthernet Connection Frequently Asked Questions This page is Intentionally Left Blank Rack Mount Slides Low Flow AlarmOptional Hardware and Software This page is Intentionally Left Blank Operation Front Panel OverviewOzone Concentration Display Menu Navigation KeypadConcentration Alarm Status/Reset Alarm Status LED Color Alarm StateStatus LED’s Status LED Normal StateEditing Numeric Values Front Panel MenusMenu Navigation Basics Front Panel Menu Diagram Front Panel Menu Diagram Diag Menu Point 5V Output MA OutputSignal Name Rear Panel Label Function Zero CalibrationSpan Calibration View Menu Parameter Meaning Units Normal RangeRange SETUP-VARS MenuMeaning Default Value Property Meaning Default Range Value SETUP-ALARM MenuConfig Menu This page is Intentionally Left Blank Digital Communications Overview Changing COM port Hardware ConfigurationDigital Communications RS232 Configuration Property Value RS-232 ConfigurationFunction DB-9 General I/O Connector Pin SW2 DTE SW2 DCE RS-485 Configuration Function DB-9 General I/O Connector Pin SW2 DCEEthernet Configuration Hardware Setup for Configuring the Ethernet ModuleVerifying Network Hardware Connection Link LED Left Side Activity LED Right Side Color MeaningDetermining a Dynamically Assigned IP Address Dhcp Mode Manually Assigning a Static IP AddressPage Instrument Setup and Configuration Registers Read / Write Modbus Register MapsReal-Time Concentration and Status Registers Read Only Decimal Star Address Description Register Type Bit BitsMaintenance Item Recommended Interval Section MaintenanceMaintenance Schedule Instrument Layout, Rack Mount Configuration Instrument LayoutInstrument Layout, Nema Configuration Replacing Internal Particulate Filter Internal Particulate Filter Replacement 1150 mV UV Lamp AdjustmentUV Lamp Replacement Sample Pump Replacement Auto-Zero Valve ReplacementConfiguring the Analog Output Cleaning Exterior Surfaces of the M465M ZeroBoot Loader Software Setup Boot Loader OperationBoot Loader Hardware Setup Upload Firmware Page Reference Drawings TroubleshootingPneumatic Diagram Interconnect Diagram Pneumatic Diagram Nema Version with Sample ConditionerInterconnect Diagram Status LED/Output Definitions Pneumatic Error Sensor OKInvalid Reading Check LampPressure Too High Calibration Factory CalibrationThis page is Intentionally Left Blank Technical Assistance This page is Intentionally Left Blank Spare Parts List for Rack Mount Version 13. M465M Spare PartsSpare Parts List for Nema Version