10.2.Troubleshooting Using Front Panel Status LED’s or Status Outputs
The M465M has 4 
Table 10-1:  Status LED/Output Definitions
| Status | Status LED | Normal | Triggers | Critical | |
| Output # | Label | Operating | 
 | 
 | Warning? | 
| 
 | 
 | State | 
 | 
 | 
 | 
| 1 | Sensor OK | On |  | Reference < 125.0 mV | Yes | 
| 
 | 
 | 
 |  Reference > 1230.0 mV – 0.5 Wt% | 
 | |
| 2 | Invalid Reading | Off |  | O3 Concentration < F.S. Range | No | 
| 
 | 
 | 
 |  | O3 Concentration > F.S. Range | 
 | 
| 3 | Check Lamp | Off |  | Reference < 250.0 mV | No | 
| 4 | Pneumatic | Off |  | Pressure < 9.0 psia | No | 
| 
 | Error | 
 |  Pressure > 18.0 psia | 
 | |
| 
 | 
 | 
 |  | Flow < 500 cc/min* | 
 | 
| *If Flow | Switch installed | 
 | 
 | 
 | 
 | 
P/N 06160D DCN6409 Teledyne API Model 465M O3 Monitor Instruction Manual - Page 66