Addendum to | Relay PCA 04523 |
2.1.1.2. AC Configuration – Standard Heaters (JP2)
Power configuration for the AC the standard heaters is set using Jumper set JP2 (see Figure
Table 2-2: Power Configuration for Standard AC Heaters (JP2)
COLOR
USED ON1
BETWEEN
PINS
FUNCTION
|
| Reaction Cell / Sample | M100E’s & | 1 to 8 | Common | |
|
| Chamber Heaters | M200E’s | 2 to 7 |
| |
|
| Neutral to Load | ||||
|
|
|
| |||
|
|
|
| 3 to 10 |
| |
|
| Mini | M200EH | Common | ||
110 VAC / 115 VAC | WHITE | Converter | 4 to 9 | Neutral to Load | ||
| ||||||
50Hz & 60 Hz |
|
| ||||
| Moly Converter | M101E and | 3 to 10 | Common | ||
|
| |||||
|
| M200E’s | 4 to 9 | Neutral to Load | ||
|
|
| ||||
|
|
|
| |||
|
|
|
| 5 to 12 |
| |
|
| By Pas Manifold | M200#M & | Common | ||
|
| M200EH | 6 to 11 | Neutral to Load | ||
|
|
| ||||
|
|
|
| |||
|
|
|
|
|
| |
|
|
|
|
|
| |
|
| Reaction Cell / Sample | M100E’s & | 1 to 7 | Load | |
|
| Chamber Heaters2 | M200E’s | |||
|
|
| ||||
220 VAC / 240 VAC |
| Hi Concentration | M100EH | 3 to 9 | Load | |
BLUE | Converter | |||||
|
|
| ||||
50Hz & 60 Hz |
|
|
|
| ||
Moly Converter | M101E and | 3 to 9 | Load | |||
| ||||||
|
| |||||
|
| M200E’s | ||||
|
|
|
|
| ||
|
|
|
|
|
| |
|
| By Pas Manifold | M200#M & | 5 to 11 | Load | |
|
| M200EH | ||||
|
|
|
|
| ||
|
|
|
|
|
|
1No AC Heaters are used in the M400E series analyzers.
Reaction Cell or Sample Chamber Heaters
Mini200EM/EH
By Pass Manifold
Heater1 | 7 | 1 | 7 |
| Reaction Cell or |
2 | 8 | 2 | 8 |
| Sample Chamber |
| Heaters | ||||
3 | 9 | 3 |
| 9 | Mini |
4 | 10 | 4 |
| 10 | Moly Converter |
| Heaters | ||||
5 | 11 | 5 | 11 |
| 200EM/EH |
|
|
|
|
| By Pass Manifold |
6 | 12 | 6 | 12 |
| Heater |
110 VAC /115 VAC | 220 VAC / 240 VAC |
|
|
Figure 2-4: Typical Set Up of AC Heater Jumper Set (JP2)
8 | 05118 Rev B3 |