Teledyne M200E and M400E DC Power Supply Test Points, DC Power Test Point and Wiring Color Code

Page 20
2.5. DC Power Supply Test Points

Addendum to E-Series Operator Manuals

Relay PCA 04523

2.5. DC Power Supply Test Points

The tables defining the names and functions of the DC power test point that appear in Chapter 11 of some revisions of the M100E/M200E/M400E operator’s manual are incorrect. The following tables show the correct information.

 

Table 2-5:

DC Power Test Point and Wiring Color Code

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NAME

 

TEST POINT#

 

COLOR

 

 

 

DEFINITION

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DGND

 

1

 

 

Black

 

 

 

Digital ground

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+5V

 

2

 

 

Red

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

AGND

 

3

 

 

Green

 

 

 

Analog ground

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+15V

 

4

 

 

Blue

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

-15V

 

5

 

 

Yellow

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+12R

 

6

 

 

Purple

 

 

12 V return (ground) line

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

+12V

 

7

 

 

Orange

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Table 2-6: DC Power Supply Acceptable Levels

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

CHECK RELAY BOARD TEST

 

 

 

 

 

POWER

VOLTAG

 

 

 

POINTS

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

MIN V

 

MAX V

 

 

 

FROM

 

TO

 

 

 

 

 

SUPPLY

 

E

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Point

 

Test Point

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

NAME

#

 

NAME

 

#

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS1

+5

 

DGND

1

+5

 

2

+4.80

 

+5.25

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS1

+15

 

AGND

3

+15

 

4

+13.5

 

+16.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS1

-15

 

AGND

3

 

-15V

 

5

-14.0

 

-16.0

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS1

AGND

 

AGND

3

 

DGND

 

1

-0.05

 

+0.05

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS1

Chassis

 

DGND

1

Chassis

 

N/A

-0.05

 

+0.05

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS2

+12

 

+12V Ret

6

 

+12V

 

7

+11.8

 

+12.5

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

PS2

DGND

 

+12V Ret

6

 

DGND

 

1

-0.05

 

+0.05

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

As on the previous version of the relay card, the test points are located at the top, right-hand corner of the PCA (see Figure 1-1)

USER NOTES:

14

05118 Rev B3

Image 20
Contents TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION T-API 9480 CARROLL PARK DRIVE SAN DIEGO, CA USA Toll-free Phone Phone FaxEmail api-sales@teledyne.com Relay PCA P/NTHIS PAGE INTENTIONALLY BLANK SAFETY MESSAGES subassemblies while the unit is under powerContacting Teledyne Instruments’ Customer Service Department at or Relay PCAAddendum to E-Series Operator Manuals LIST OF TABLES TABLE OF CONTENTSLIST OF FIGURES THIS PAGE INTENTIONALLY BLANK 1.1. What’s the Same Table 1-1 List of Analyzers Using New Relay PCA P/N045231. INTRODUCTION Figure 1-1 Relay PCA Layout P/N04523 1.2. What’s Different1.3. Relay PCA Location and Layout Figure 1-3 Relay PCA P/N 045230100 with Safety Shield In Place 1.3.1. Version1.3.2. Version Figure 1-2 Relay PCA P/N 045230200 with AC Relay Retainer In PlaceUSER NOTES 2. OPERATION AND CONFIGURATION 2.1. Power Distribution2.1.1. AC Power Configuration Figure 2-1 Power Distribution Block DiagramFigure 2-2 Location of AC power Configuration Jumpers IZS Permeation Tube Heater and O2 Sensor Connection optional2.1.1.1. AC configuration - Internal Pump JP7 AC Power Configuration for Internal Pumps JP7Figure 2-3 Pump AC Power Jumpers JP7 110 VAC /115 VAC2.1.1.2. AC Configuration - Standard Heaters JP2 Table 2-2 Power Configuration for Standard AC Heaters JP2Figure 2-4 Typical Set Up of AC Heater Jumper Set JP2 1 to2.1.1.3. AC Configuration -Heaters for Option Packages JP6 Table 2-3 Power Configuration for Optional AC Heaters JP6Figure 2-5 Typical Set Up of AC Heater Jumper Set JP2 2.2. Valve Control2.3.1. Watchdog Indicator D1 2.3. Status LEDs & Watch Dog CircuitryD1 RED Figure 2-6 Status LED Locations - Relay PCAFigure 2-7 Heater Control Loop Block Diagram 2.4. Heater Control LoopThemocouples MOTHER BOARD2.4.1. Thermocouple Inputs and Configuration Jumper JP5 Table 2-4 Thermocouple Configuration Jumper JP5 Pin-OutsType J Compensation Type K CompensationFigure 2-8 Thermocouple Configuration Jumper JP5 Pin-Outs INPUT TC1 J15Output Scale Selector 7 Type J Compensation 82.5. DC Power Supply Test Points DC Power Test Point and Wiring Color CodeTable 2-6 DC Power Supply Acceptable Levels FROM3. TROUBLE SHOOTING AND REPAIR 3.1. Removing / Replacing the Relay PCA from the InstrumentMounting Screws Figure 3-1 Relay PCA 04523 Mounting Screw Locations4. SCHEMATICS AND SPARE PARTS FOR RELAY PCA P/N Table 4-2 List of Included Electronic SchematicsList of Spare Parts DOCUMENT #