Teledyne M200E and M400E Thermocouple Inputs and Configuration Jumper JP5, Type J Compensation

Page 18
2.4.1. Thermocouple Inputs and Configuration Jumper (JP5)

Addendum to E-Series Operator Manuals

Relay PCA 04523

2.4.1. Thermocouple Inputs and Configuration Jumper (JP5)

Although the 04523 relay PCA supports two thermocouple inputs, the current M100E/M200E series analyzers only utilize one and the M400E series analyzers none. By default, this single thermocouple input is plugged into the TC1 input (J15). TC2 (J16) is currently not used. See Figure 1-1 for location of J15 and J16

Table 2-4: Thermocouple Configuration Jumper (JP5) Pin-Outs

TC INPUT

JUMPER PAIR

DESCRIPTION

FUNCTION

 

 

 

 

 

 

 

 

 

 

 

Selects preamp gain factor for J or K TC

 

1 – 11

Gain Selector

- IN = J TC gain factor

 

 

 

- OUT = K TC gain factor

 

 

 

Selects preamp gain factor for J or K TC

 

 

 

 

2 – 12

Output Scale Selector

- IN = 5 mV / °C

 

 

 

- OUT = 10 mV / °C

TC1

 

 

When present, sets Cold Junction

3 – 13

Type J Compensation

Compensation for J type Thermocouple

 

 

 

When present, sets Cold Junction

 

4 – 14

Type K Compensation

 

Compensation for K type Thermocouple

 

 

 

 

 

 

Selects between Isolated and grounded

 

 

 

 

5 – 15

Termination Selector

TC

 

- IN = Isolate TC

 

 

 

 

 

 

- OUT = Grounded TC

 

 

 

 

TC2

6 – 16

Gain Selector

 

 

7 – 17

Output Scale Selector

 

 

8 – 18

Type J Compensation

 

 

9 – 19

Type K Compensation

 

 

10 – 20

Termination Selector

 

 

Same as Pins 1 – 11 above. Same as Pins 2 – 12 above. Same as Pins 3 – 13 above. Same as Pins 4 – 14 above. Same as Pins 5 – 15 above.

12

05118 Rev B3

Image 18
Contents Email api-sales@teledyne.com TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION T-API9480 CARROLL PARK DRIVE SAN DIEGO, CA USA Toll-free Phone Phone Fax Relay PCA P/NTHIS PAGE INTENTIONALLY BLANK Contacting Teledyne Instruments’ Customer Service Department at or SAFETY MESSAGESsubassemblies while the unit is under power Relay PCAAddendum to E-Series Operator Manuals TABLE OF CONTENTS LIST OF FIGURESLIST OF TABLES 05118 Rev B3 Table 1-1 List of Analyzers Using New Relay PCA P/N04523 1. INTRODUCTION1.1. What’s the Same 1.2. What’s Different 1.3. Relay PCA Location and LayoutFigure 1-1 Relay PCA Layout P/N04523 1.3.2. Version Figure 1-3 Relay PCA P/N 045230100 with Safety Shield In Place1.3.1. Version Figure 1-2 Relay PCA P/N 045230200 with AC Relay Retainer In PlaceUSER NOTES 2.1.1. AC Power Configuration 2. OPERATION AND CONFIGURATION2.1. Power Distribution Figure 2-1 Power Distribution Block DiagramFigure 2-2 Location of AC power Configuration Jumpers IZS Permeation Tube Heater and O2 Sensor Connection optionalFigure 2-3 Pump AC Power Jumpers JP7 2.1.1.1. AC configuration - Internal Pump JP7AC Power Configuration for Internal Pumps JP7 110 VAC /115 VACFigure 2-4 Typical Set Up of AC Heater Jumper Set JP2 2.1.1.2. AC Configuration - Standard Heaters JP2Table 2-2 Power Configuration for Standard AC Heaters JP2 1 toFigure 2-5 Typical Set Up of AC Heater Jumper Set JP2 2.1.1.3. AC Configuration -Heaters for Option Packages JP6Table 2-3 Power Configuration for Optional AC Heaters JP6 2.2. Valve ControlD1 RED 2.3.1. Watchdog Indicator D12.3. Status LEDs & Watch Dog Circuitry Figure 2-6 Status LED Locations - Relay PCAThemocouples Figure 2-7 Heater Control Loop Block Diagram2.4. Heater Control Loop MOTHER BOARDType J Compensation 2.4.1. Thermocouple Inputs and Configuration Jumper JP5Table 2-4 Thermocouple Configuration Jumper JP5 Pin-Outs Type K CompensationOutput Scale Selector 7 Figure 2-8 Thermocouple Configuration Jumper JP5 Pin-OutsINPUT TC1 J15 Type J Compensation 8Table 2-6 DC Power Supply Acceptable Levels 2.5. DC Power Supply Test PointsDC Power Test Point and Wiring Color Code FROMMounting Screws 3. TROUBLE SHOOTING AND REPAIR3.1. Removing / Replacing the Relay PCA from the Instrument Figure 3-1 Relay PCA 04523 Mounting Screw LocationsList of Spare Parts 4. SCHEMATICS AND SPARE PARTS FOR RELAY PCA P/NTable 4-2 List of Included Electronic Schematics DOCUMENT #