Relay PCA 04523 | Addendum to |
2.1.1.1. AC configuration – Internal Pump (JP7)
AC power configuration for internal pumps is set using Jumper set JP7 (see Figure
|
| Table | AC Power Configuration for Internal Pumps (JP7) |
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
| LINE | LINE | JUMPER |
| JUMPER |
|
| FUNCTION | BETWEEN |
| |||
| POWER | FREQUENCY | COLOR |
| ||
|
| PINS |
| |||
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
| Connects pump pin 3 to 110 / 115 VAC power line | 2 to 7 |
|
|
| 60 HZ | WHITE |
|
|
|
|
| Connects pump pin 3 to 110 / 115 VAC power line | 3 to 8 |
| ||
|
|
|
|
|
|
|
| 110VAC |
|
| Connects pump pins 2 & 4 to Neutral | 4 to 9 |
|
| 115 VAC |
|
| Connects pump pin 3 to 110 / 115 VAC power line | 2 to 7 |
|
|
|
|
|
| ||
|
| 50 HZ1 | BLACK |
|
|
|
|
| Connects pump pin 3 to 110 / 115 VAC power line | 3 to 8 |
| ||
|
|
|
| Connects pump pins 2 & 4 to Neutral | 4 to 9 |
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
| 60 HZ | BROWN | Connects pump pins 3 and 4 together | 1 to 6 |
|
| 220VAC |
|
|
| ||
| Connects pump pin 1 to 220 / 240VAC power line | 3 to 8 |
| |||
|
|
|
| |||
| 240 VAC | 50 HZ1 | BLUE | Connects pump pins 3 and 4 together | 1 to 6 |
|
|
|
|
|
| ||
|
| Connects pump pin 1 to 220 / 240VAC power line | 3 to 8 |
| ||
|
|
|
|
| ||
|
|
|
|
|
|
|
1 A jumper between pins 5 and 10 may be present on the jumper plug assembly, but is only functional on the M300E and has no function on the Models M100E, M200E or M400E.
110 VAC /115 VAC |
| 220 VAC /240 VAC | ||
1 | 6 |
| 1 | 6 |
2 | 7 |
| 2 | 7 |
3 |
| 8 | 3 | 8 |
4 | 9 |
| 4 | 9 |
5 | 10 |
| 5 | 10 |
Present on 50 Hz version of jumper set,
and functional for M300E but not Models M100E, M200E & M400E
Figure 2-3: Pump AC Power Jumpers (JP7)
05118 Rev B3 | 7 |