Read Flow

Selecting Read flow informs the user of the mode of operation and the approximate maximum flow rate. Press the appropriate key can change the units required.

Attach sensor set in REFLEX mode Approx. max. flow:

7.22m/s

ENTER to continue

SCROLL changes mode

Pressing ENTER asks the user to enter a temperature in °C.

Fluid temp?

20.0

(°C)

 

Now press scroll (up). The instrument will display the separation distance before displaying flow.

Main Menu - Select Sensor mode

When the application information is programmed into the instrument it selects and defaults to the most suitable mode of operation i.e. REFLEX or DIAGONAL.

Sensor Mode

Selecting Sensor mode allows the user to choose the appropriate method for clamping the sensors to the pipe. The default would have been displayed on the previous screen and Sensor mode can be selected to give the user a choice between Reflex and Diagonal.

SENSOR SET

 

Mode

REFLEX

Read flow

 

Exit and default

 

This option is available for two main reasons. Firstly, lets assume that the instrument has selected “mount sensors in DIAGONAL MODE” Your application may not allow you to achieve this mode. Providing the velocity is low enough it is possible to force the sensors into REFLEX mode (See page 4). Changing the sensor mode from Diagonal to Reflex would allow the user to measure the flow. The display may also read, sensor mode invalid for this pipe size.

Cannot READ FLOW Because pipe

to large/small for sensor

ENTER to continue

Read Flow

Moving the cursor to Read flow and pressing ENTER informs the user of the mode of operation and the maximum flow capable.

Should the actual flow be higher than the one specified on the instrument, another mode of operation can be selected. Selecting EXIT will take you back to MAIN MENU.

11

Image 11
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Pulse Cable Connections ConnectorsTransducers Battery CircuitKeypad Main Menu Quick Start Switch 0n…Option number for the software Main Menu Millimetres EnterSelect Enter and the display will now show Attach SensorsVIEW/EDIT Site Data Error Messages Appear HereMain Menu View/Edit Site Data XXXSite Name Dimension UnitsList Sites Site NumberSensor SET Main Menu Select Sensor modeSensor Mode Read FlowApplication Options Main Menu Set Up Instrument Pulse Output KeyEnabled Display backlightOptions Key Main Menu Read FlowDelete Key Calibration Factor Reset + Total/- TotalTotal Damping secStatus Messages Error MessagesOther Messages Temperature range is -20C to +125CW6 mA OUT Overange W8 mA Load to HighEnter a lining thickness first Reflex modeMounting the Transducers Transducer PositioningReynolds Number CorrectincorrectLiquid Conditions Propagation Velocity or Sound SpeedMaximum Flow Application TemperatureTo calculate Re use the following formula Gull wing and spring attachment Peizo electric crystal Red connector upstreamBlue connector downstream Outputs Battery PackEnclosure Supply VoltageWarranty H2O Liquid Sound SpeedsSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C2H5I C4H9NOCH2I2 C2H3N CH3CO2O C2H6O C2H7NOC7H9N C7H9N CH4O C7H8 C6H5FC3H8O C3H8O CH4 CH4OC3H8O C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2OKNO3 H2O CH4O C2H6O H2SO4CF4 C4H9NO Software Charging Controller ICCharging Voltage Instrument differencesExamples Quicker full charge