Warning Messages

W1: CHECK SITE DATA

This message occurs when the application information has been entered incorrectly and the wrong sensors have been attached to the wrong pipe size causing the system timing to be in error. The site data needs to be checked and the instrument reprogrammed.

W2: SIGNAL TIMING POOR

Unstable signal timing or differing up/down stream times indicate that the liquid is aerated or pipe surface is of poor quality.

W5: FLOW SIGNALS POOR

This warning appears when there is a signal lower than 25%. This could be due to the application, a poor quality pipe, amongst others.

W6: mA OUT OVERANGE

The mA output is over-range when the flow is higher than the maximum mA range. Once the

4-20mA has been set up and the flow goes above the range set then this message will appear. It is possible to re-scale the 4-20mA to be able to cope with the higher flow.

W7: BATTERY LOW

The battery low warning occurs when battery indication is on 40%. The instrument has approximately 30 minutes usage before it needs recharging. (See application note at the back of this manual)

W8: mA LOAD TO HIGH

The 4-20mA Output is designed to work with a load up to 750. When the load is too high or not connected, the above warning message will be displayed.

Other Messages

The messages below appear mainly when data has been incorrectly entered or the Portaflow 216 is trying to be used on an application that it is not capable of working on.

Pipe OD out of range

The outside diameter of the pipe has been entered and is out of range of the instrument.

Wall thickness out of range

The wall thickness that has been entered is out of range of the instrument.

Lining thickness out of range

The pipe lining thickness has been incorrectly entered.

Site range is 0 - 20

There are only 20 storage sites available with 0 being the QUICK START site.

CANNOT READ FLOW BECAUSE… ….Pipe dimensions are invalid

CANNOT READ FLOW BECAUSE …materials are invalid

CANNOT READ FLOW BECAUSE …Pipe is too large for sensor set

CANNOT READ FLOW BECAUSE …Pipe is too small for sensor set

CANNOT READ FLOW BECAUSE

…Sensor mode is invalid for this pipe size

Temperature range is -20°C to +125°C

The temperature range of the transducers is -20°C to +125°C.

16

Image 16
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Connectors Pulse Cable ConnectionsKeypad Battery CircuitTransducers Option number for the software Main Menu Switch 0n…Main Menu Quick Start Enter MillimetresAttach Sensors Select Enter and the display will now showError Messages Appear Here Main Menu View/Edit Site DataXXX VIEW/EDIT Site DataDimension Units List SitesSite Number Site NameMain Menu Select Sensor mode Sensor ModeRead Flow Sensor SETMain Menu Set Up Instrument Pulse Output Key EnabledDisplay backlight Application OptionsDelete Key Main Menu Read FlowOptions Key Reset + Total/- Total TotalDamping sec Calibration FactorError Messages Status MessagesTemperature range is -20C to +125C W6 mA OUT OverangeW8 mA Load to High Other MessagesReflex mode Enter a lining thickness firstTransducer Positioning Mounting the TransducersCorrectincorrect Liquid ConditionsPropagation Velocity or Sound Speed Reynolds NumberTo calculate Re use the following formula Application TemperatureMaximum Flow Blue connector downstream Red connector upstreamGull wing and spring attachment Peizo electric crystal Battery Pack EnclosureSupply Voltage OutputsWarranty Liquid Sound Speeds H2OSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C4H9NO CH2I2C2H3N CH3CO2O C2H6O C2H7NO C2H5IC6H5F C3H8O C3H8OCH4 CH4O C7H9N C7H9N CH4O C7H8KNO3 C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2OC3H8O CF4 C4H9NO C2H6O H2SO4H2O CH4O Charging Controller IC Charging VoltageInstrument differences SoftwareQuicker full charge Examples