Micronics 216 To calculate Re use the following formula, Maximum Flow, Application Temperature

Page 20

If the Portaflow 216 is to be used on laminar flow applications it will be necessary to calculate the Reynolds No for each application. To calculate the Reynolds No it is necessary to know the Kinematic viscosity in Centistokes; the flow velocity and the pipe inside diameter. Please follow the table below

To calculate Re use the following formula: -

Re =

dv

(7730)

or

Re =

d 1

v1

(1000)

υ

1

υ

1

 

 

 

 

 

Where

d= inside pipe diameter in inches

d 1 = inside pipe diameter in millimetres

v= velocity in feet/second

v1 = velocity in metres/second

υ1 = Kinematic viscosity in centistokes

MAXIMUM FLOW

The maximum flow is dependent on the velocity and pipe size.

APPLICATION TEMPERATURE

On any application whose operating temperature is either above or below ambient temperature ensure that the transducers reach and are maintained at the application temperature before undertaking a measurement.

When applying the transducers to low temperature applications do not allow the pipe surface to ice up between the transducer and the pipe wall. The ice will force the block away from the pipe wall and consequently you will lose the signal.

DIAGONAL MODE SETUP

Figure 11: Diagonal Mode parts supplied

Gull-wing attachment

Spring tensioning

Sensor stem

Piezo electric crystal

As part of your New Portaflow 216 kit you will find two stainless steel gull wings, two springs and two lengths of chain.

Take the transducers from the reflex guiderail. Attach the Gull-wing to each transducer using the washer & wing nut provided.

Apply grease to the bottom of the transducer (as shown on page3). Wrap the chain around the pipe as shown. Expand the spring and carefully slide the chain into the slot on the Gull Wing. Plug the red connector into the socket on the upstream sensor. The sensor with the red cable must be positioned up stream. The stem of the sensor must point towards the downstream sensor.

20

Image 20
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Connectors Pulse Cable ConnectionsTransducers Battery CircuitKeypad Main Menu Quick Start Switch 0n…Option number for the software Main Menu Enter MillimetresAttach Sensors Select Enter and the display will now showError Messages Appear Here Main Menu View/Edit Site DataXXX VIEW/EDIT Site DataDimension Units List SitesSite Number Site NameMain Menu Select Sensor mode Sensor ModeRead Flow Sensor SETMain Menu Set Up Instrument Pulse Output Key EnabledDisplay backlight Application OptionsOptions Key Main Menu Read FlowDelete Key Reset + Total/- Total TotalDamping sec Calibration FactorError Messages Status MessagesTemperature range is -20C to +125C W6 mA OUT OverangeW8 mA Load to High Other MessagesReflex mode Enter a lining thickness firstTransducer Positioning Mounting the TransducersCorrectincorrect Liquid ConditionsPropagation Velocity or Sound Speed Reynolds NumberMaximum Flow Application TemperatureTo calculate Re use the following formula Gull wing and spring attachment Peizo electric crystal Red connector upstreamBlue connector downstream Battery Pack EnclosureSupply Voltage OutputsWarranty Liquid Sound Speeds H2OSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C4H9NO CH2I2C2H3N CH3CO2O C2H6O C2H7NO C2H5IC6H5F C3H8O C3H8OCH4 CH4O C7H9N C7H9N CH4O C7H8C3H8O C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2OKNO3 H2O CH4O C2H6O H2SO4CF4 C4H9NO Charging Controller IC Charging VoltageInstrument differences SoftwareQuicker full charge Examples