Micronics 216 manual Enter, Millimetres

Page 7

The instrument now asks for the Pipe outside diameter? After entering the outside diameter in millimeters press

ENTER.

QUICK START

 

Dimension units

mm

Pipe O.D.?

58.0

Pipe wall thickness now appears on the display. Enter the pipe wall thickness in millimeters, then press ENTER.

QUICK START

 

Dimension units

MILLIMETRES

Pipe O.D.?

58.0

Wall thick?

4.0

 

 

Pipe lining thickness now appears on the display. If the pipe you are measuring has a lining, enter the Pipe lining thickness. If nothing is entered the instrument automatically assumes there is no lining. Press ENTER to move onr after entering the data.

QUICK START

 

Dimension units

MILLIMETRES

Pipe outside diameter?

58.0

Wall thick?

4.0

Lining?

0.0

 

 

The instrument now displays Select pipe wall material. Using the scroll keys it is possible to scroll up or down the options available. Select the required material and press ENTER.

QUICK START

Select pipe wall material:

Mild Steel

S’ less Steel 316

S’ less Steel 303

Plastic

Cast Iron

Ductile Iron

Copper

Brass

Concrete

Glass

Other (m/s)

The following will only be displayed at this stage if a lining thickness has been entered. Use the scroll keys to select the required material, then press ENTER. If Other is selected, enter the sound speed of the lining in metres/sec.

7

Image 7
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Pulse Cable Connections ConnectorsKeypad Battery CircuitTransducers Option number for the software Main Menu Switch 0n…Main Menu Quick Start Millimetres EnterSelect Enter and the display will now show Attach SensorsVIEW/EDIT Site Data Error Messages Appear HereMain Menu View/Edit Site Data XXXSite Name Dimension UnitsList Sites Site NumberSensor SET Main Menu Select Sensor modeSensor Mode Read FlowApplication Options Main Menu Set Up Instrument Pulse Output KeyEnabled Display backlightDelete Key Main Menu Read FlowOptions Key Calibration Factor Reset + Total/- TotalTotal Damping secStatus Messages Error MessagesOther Messages Temperature range is -20C to +125CW6 mA OUT Overange W8 mA Load to HighEnter a lining thickness first Reflex modeMounting the Transducers Transducer PositioningReynolds Number CorrectincorrectLiquid Conditions Propagation Velocity or Sound SpeedTo calculate Re use the following formula Application TemperatureMaximum Flow Blue connector downstream Red connector upstreamGull wing and spring attachment Peizo electric crystal Outputs Battery PackEnclosure Supply VoltageWarranty H2O Liquid Sound SpeedsSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C2H5I C4H9NOCH2I2 C2H3N CH3CO2O C2H6O C2H7NOC7H9N C7H9N CH4O C7H8 C6H5FC3H8O C3H8O CH4 CH4OKNO3 C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2OC3H8O CF4 C4H9NO C2H6O H2SO4H2O CH4O Software Charging Controller ICCharging Voltage Instrument differencesExamples Quicker full charge