Micronics 216 manual Main Menu Set Up Instrument Pulse Output Key, Display backlight, Enabled

Page 12

Main Menu - Set Up Instrument

Pulse Output Key

This can only be operated in flow mode.

Use the scroll key to move the cursor up or down the display. To change the flow units press the key required. This will also change the flow units when returning to the flow mode. Changing the flow units will also re-scale the litres per pulse.

PULSE OUTPUT

Flow units

Output OFF

Max. pulse rate1 per sec

Litres per pulse12.76

Exit

Outputs allow the user to select from the following.

Selecting Off switches the pulse off and returns to the PULSE OUTPUT display.

Selecting the Forward total counts the pulses of the forward flow only.

Selecting Net total counts the pulses of the sum of the forward total less the reverse total.

OUTPUT

Off

Forward total

Net total

Max. Pulse Rate

This option allows the user to select between fast/slow pulses or large/small pulse width. Select 1 per second for slow pulses and 100 for a fast pulse. The pulse width for 1 per second is 100ms and 5ms for 100 per second.

XXXX per pulse

This will change when the flow units are changed above. When the correct flow units are selected this allows the user to scale the pulses to there own requirements or it can be left in the default setting.

Display backlight

Use the scroll key to select backlight and press ENTER. This allows the user to enable or disable the backlight. Enable, means the backlight will stay on for 15secs with every key press. It will stay on permanently with the mains plugged in. Use the scroll key to select and press ENTER. The backlight will draw power from the batteries and reduce the operating life of the battery cell. (follow the application note at the back of the manual)

Backlight

Enabled

Disabled

Application Options

Use the scroll key to select Application Options and press ENTER. Please enter the following password 39502600. It is a facility that could enhance signal levels on difficult applications, primarily very small or very large pipes. Use enhanced mode when signals are below 800. Below 800 the system may generate noise and therefore the accuracy of the measurement cannot be guaranteed.

12

Image 12
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Connectors Pulse Cable ConnectionsBattery Circuit KeypadTransducers Switch 0n… Option number for the software Main MenuMain Menu Quick Start Enter MillimetresAttach Sensors Select Enter and the display will now showError Messages Appear Here Main Menu View/Edit Site DataXXX VIEW/EDIT Site DataDimension Units List SitesSite Number Site NameMain Menu Select Sensor mode Sensor ModeRead Flow Sensor SETMain Menu Set Up Instrument Pulse Output Key EnabledDisplay backlight Application OptionsMain Menu Read Flow Delete KeyOptions Key Reset + Total/- Total TotalDamping sec Calibration FactorError Messages Status MessagesTemperature range is -20C to +125C W6 mA OUT OverangeW8 mA Load to High Other MessagesReflex mode Enter a lining thickness firstTransducer Positioning Mounting the TransducersCorrectincorrect Liquid ConditionsPropagation Velocity or Sound Speed Reynolds NumberApplication Temperature To calculate Re use the following formulaMaximum Flow Red connector upstream Blue connector downstreamGull wing and spring attachment Peizo electric crystal Battery Pack EnclosureSupply Voltage OutputsWarranty Liquid Sound Speeds H2OSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C4H9NO CH2I2C2H3N CH3CO2O C2H6O C2H7NO C2H5IC6H5F C3H8O C3H8OCH4 CH4O C7H9N C7H9N CH4O C7H8C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2O KNO3C3H8O C2H6O H2SO4 CF4 C4H9NOH2O CH4O Charging Controller IC Charging VoltageInstrument differences SoftwareQuicker full charge Examples