Micronics 216 manual Inverted view of grease applied to sensor block

Page 3

Select Quick Start - Press ENTER.

Dimension Units? – Scroll to select units required, press ENTER.

Pipe OD – Enter data, press ENTER.

Pipe Wall Thickness – Enter data, press ENTER.

Pipe Lining Thickness – Enter data, press ENTER. ENTER Zero if there is no lining on the application. Select Wall Material – Select using scroll keys, press ENTER.

Select Lining Material – This will only be displayed if a lining thickness has been entered. Select using the scroll keys, press ENTER.

Select Fluid Type – Select using scroll keys, press ENTER.

The instrument selects the mode of operation using the data entered and will display the following.

Attach sensor set in XXXX mode Approx. max. flow:

XXX m/s

Press ENTER to continue Or SCROLL to change mode

Fluid Temp? Press Enter to Input the application temperature in the units required °C or °F then press ENTER.

Now retract the sensor blocks back into the guide rail by turning the locking nuts clockwise.

Apply grease to both sensor blocks as shown in (Figure 1), attach to the pipe using the appropriate mounting hardware in either Reflex or Diagonal Mode. Ensure the Guide rail itself is free of grease

Figure 1:- Inverted view of grease applied to sensor block

Connect the red and blue sensor cables to the electronics and the guide rail assembly. The RED cable must be positioned upstream to give a positive flow reading.

Figure 2:- Sensor assembly

3

Image 3
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Pulse Cable Connections ConnectorsBattery Circuit KeypadTransducers Switch 0n… Option number for the software Main MenuMain Menu Quick Start Millimetres EnterSelect Enter and the display will now show Attach SensorsVIEW/EDIT Site Data Error Messages Appear HereMain Menu View/Edit Site Data XXXSite Name Dimension UnitsList Sites Site NumberSensor SET Main Menu Select Sensor modeSensor Mode Read FlowApplication Options Main Menu Set Up Instrument Pulse Output KeyEnabled Display backlightMain Menu Read Flow Delete KeyOptions Key Calibration Factor Reset + Total/- TotalTotal Damping secStatus Messages Error MessagesOther Messages Temperature range is -20C to +125CW6 mA OUT Overange W8 mA Load to HighEnter a lining thickness first Reflex modeMounting the Transducers Transducer PositioningReynolds Number CorrectincorrectLiquid Conditions Propagation Velocity or Sound SpeedApplication Temperature To calculate Re use the following formulaMaximum Flow Red connector upstream Blue connector downstreamGull wing and spring attachment Peizo electric crystal Outputs Battery PackEnclosure Supply VoltageWarranty H2O Liquid Sound SpeedsSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C2H5I C4H9NOCH2I2 C2H3N CH3CO2O C2H6O C2H7NOC7H9N C7H9N CH4O C7H8 C6H5FC3H8O C3H8O CH4 CH4OC7H8O C4H9NO C6H5NO2 CH3NO2 C9H2O KNO3C3H8O C2H6O H2SO4 CF4 C4H9NOH2O CH4O Software Charging Controller ICCharging Voltage Instrument differencesExamples Quicker full charge