Micronics 216 manual Reset + Total/- Total, Damping sec, Calibration Factor, Correction Factor

Page 14

Set Zero Flow

On some applications and in some conditions it may be that although there is no flow the instrument may show a small offset due to system noise. The offset can be cancelled out and will increase the accuracy of the instrument. Selecting this option and pressing ENTER the display will show the following.

Stop the flow COMPLETELY and press ENTER or SCROLL to cancel

Pressing ENTER before the flow has stopped will result in an error message asking if you are you sure the flow has stopped. This occurs when the flow is still above 0.25m/sec.

When this facility has already been selected, press ENTER to cancel the previous instruction, then it is possible to re- set the Zero balance. The option is not available when error messages E1 and E2 (See page 17) are being displayed.

Total

This option allows the user to disable the positive and negative totalizers. When you select either of these options the totaliser will start or stop functioning. It does not zero the total, this is a separate function described below.

Reset + Total/- Total

The Portaflow 216 has forward and reverse totaliser that can be reset when this facility is selected. Use the scroll keys to select and press ENTER to reset. The Total is stored when the unit is switched off or battery goes flat, therefore may need to be reset before each use.

Damping (sec)

This facility is used when the flow readings are unstable due to turbulence caused by obstructions or bends etc. Damping or averaging can be used to make the readings more stable. It can be set to up-date the display, anything between 3 and 100 seconds.

Calibration Factor

If for any reason the instrument goes out of calibration and the readings may be higher or lower than normal then this facility enables the user to correct the reading. If for example the reading is 4% higher than normal then entering 0.96 will reduce the reading by 4%. If the reading is 4% lower than normal then entering 1.04 would increase the reading by 4%.

When the instrument is supplied it will always default to 1.00 and when this is changed it will stay in the memory to whatever it has been changed to, until such time as it needs to be changed again.

Correction Factor

This is a facility that can be used when errors occur due to lack of straight pipe or the sensors have been placed too close to a bend, this could give an incorrect reading to what is expected. The user can set this as a % in the same way as the calibration factor, but it will not be stored in the memory.

Diagnostics

Calculated s

This is a value the instrument predicts will be the time in ∝ secs that it should take for the transmitted signal to go across a particular pipe size. This value is ascertained from the data entered by the user. i.e. Pipe size, material, sensor set etc.

Up s, Dn s

This is the actual transit time measured by the instrument and will be slightly (5-10∝ s depending on the pipe size and signal condition) less than the calculated value above.

14

Image 14
Contents Portaflow Fast Track Set up Procedure Inverted view of grease applied to sensor block Connectors Pulse Cable ConnectionsTransducers Battery CircuitKeypad Main Menu Quick Start Switch 0n…Option number for the software Main Menu Enter MillimetresAttach Sensors Select Enter and the display will now showXXX Error Messages Appear HereMain Menu View/Edit Site Data VIEW/EDIT Site DataSite Number Dimension UnitsList Sites Site NameRead Flow Main Menu Select Sensor modeSensor Mode Sensor SETDisplay backlight Main Menu Set Up Instrument Pulse Output KeyEnabled Application OptionsOptions Key Main Menu Read FlowDelete Key Damping sec Reset + Total/- TotalTotal Calibration FactorError Messages Status MessagesW8 mA Load to High Temperature range is -20C to +125CW6 mA OUT Overange Other MessagesReflex mode Enter a lining thickness firstTransducer Positioning Mounting the TransducersPropagation Velocity or Sound Speed CorrectincorrectLiquid Conditions Reynolds NumberMaximum Flow Application TemperatureTo calculate Re use the following formula Gull wing and spring attachment Peizo electric crystal Red connector upstreamBlue connector downstream Supply Voltage Battery PackEnclosure OutputsWarranty Liquid Sound Speeds H2OSubstance Form Index Specific Gravity Sound Speed ∆ v/ºC C2H3N CH3CO2O C2H6O C2H7NO C4H9NOCH2I2 C2H5ICH4 CH4O C6H5FC3H8O C3H8O C7H9N C7H9N CH4O C7H8C3H8O C7H8O C4H9NO C6H5NO2 CH3NO2 C9H2OKNO3 H2O CH4O C2H6O H2SO4CF4 C4H9NO Instrument differences Charging Controller ICCharging Voltage SoftwareQuicker full charge Examples