Step 3. Set the Logical Address for each M-Module

Guidelines for Determining Logical Addresses

Each VXI module or M-Module in a VXI system must have a unique logical address. The logical address is used by the VXI resource manager to identify instruments. When the logical address of a module is set to a direct multiple of 8 (such as 8, 16, 24, ... 256), that module is considered an “instrument”.

An instrument's secondary address (used to program the module from GPIB) is derived by dividing the instrument logical address by 8. For example, a logical address of 24 is a secondary address of 03.

The E2251A comes from the factory with the M-Module logical address

switches set to 0.1 In the E2251A M-Module Carrier, logical address 0 instructs the Carrier to ignore that slot and the Carrier does not report it to the VXI system resource manager.

You must set the logical address switches for each M-Module installed in the E2251A Carrier.

If you are using an E1406 Command Module set M-Module addresses to a multiple of 8 (i.e., 8, 16, 24, 32, . . . 256). Do not set any M-Module logical address to the same value as any other M-Module or any other VXI module in the system. You can use up to 31 M-Modules (up to seven Carriers) with an E1406 Command Module or 255 M-Modules (up to 51 Carriers) with embedded controllers, VXLink, or MXIbus controllers.

Note

Unused Slots on the Carrier

Multiple-module Scanning Voltmeter or Switchbox instruments are not supported with M-Modules on the E2251A Carrier. Multiple-module instruments can only be created using standard B- or C-size VXI modules.

Unused slots on the E2251A Carrier should have their logical address set to 0 (zero). With the logical address set to 0, the Carrier disables the circuits (interrupt, reserved memory, etc.) for that slot.

1.Note that in a standard VXI system, logical address 0 is reserved for the VXIbus command module / resource manager. On the E2251A Carrier however, logical address 0 causes the Carrier to disregard that M-Module slot.

12

Image 14
Contents E2251-90002 Page Contents Contents Agilent Technologies Warranty Statement Documentation History Page Declaration of Conformity Page Page Module Installation Remove the Carrier Shields Determine M-Module Layout Set the Logical Address for each M-Module Guidelines for Determining Logical AddressesModule M3 Module M2 Module M1 Install M-Modules onto the Carrier Connect Field Wiring to Slot M4 & M5 M-Modules Reattach the Carrier Shields Install Carrier into C-Size VXI Mainframe Connect Field Wiring to Front M-Modules Installation / Operational Verification Removing M-Modules from the Carrier Module Field Wiring Information Assembling the M-Module Field Wiring ConnectorPage Place connector Connector screws on bottom shell E2259A Double-Wide Breadboard M-Module Wiring Information E2259A J102 Pinout Table E2261A Quad RS-232 Interface Wiring Information Serial Port 1 Serial Port Serial Port 3 Serial PortPage E2270A 16-Channel Form a Switch Wiring Information E2271A 4x4 Matrix Switch Wiring Information ROW2LOChannel Configuring the E2272A E2273A 8-Channel Form C Switch Wiring Information E2274A 4-Channel Power Relay Wiring Information E2290A 16-Bit Digital I/O Wiring Information UseE2291A 16-Channel Isolated Digital Output Wiring Information Page Index NumericsPosition in Carrier, 11 removing from Carrier, 20 slots