Agilent Technologies E2251A user manual E2270A 16-Channel Form a Switch Wiring Information

Page 30

E2270A 16-Channel Form A Switch Wiring Information

 

 

MAXIMUM VOLTAGE/CURRENT: The maximum voltage

 

 

that may be applied to any connector on the E2270A,

 

 

E2271A, or E2272A is 200 VDC, 125 VAC rms, or 175 VAC

 

 

peak. These limits apply only if the product is installed

 

 

in a humidity-controlled (<60% RH) environment where

 

 

airborne contaminants and transients are controlled,

 

 

and there is NOT a relay connection made to power

 

 

mains. If these conditions CANNOT be maintained, then

 

 

the maximum voltage is 60 VDC, 48 VAC rms or 68 VAC

 

 

peak.

 

 

 

The maximum current (non-inductive) that may be ap-

 

 

plied to the E2270A, E2271A, or E2272A is:

 

 

Per Switch: 2 ADC, 2 AAC peak

 

 

 

Per M-Module: 8ADC, 8AAC peak

 

User

 

User

 

Connector

Connector

 

Pins

 

Pins

 

Channel 0

 

Channel 8

 

1

CH0_NO

9

CH8_NO

16

CH0_COM

24

CH8_COM

Channel 1

 

Channel 9

 

2

CH1_NO

10

CH9_NO

17

CH1_COM

25

CH9_COM

Channel 2

 

Channel 10

 

3

CH2_NO

11

CH10_NO

18

CH2_COM

26

CH10_COM

Channel 3

 

Channel 11

 

4

CH3_NO

12

CH11_NO

19

CH3_COM

27

CH11_COM

Channel 4

 

Channel 12

 

5

CH4_NO

13

CH12_NO

20

CH4_COM

28

CH12_COM

Channel 5

 

Channel 13

 

6

CH5_NO

14

CH13_NO

21

CH5_COM

29

CH13_COM

Channel 6

 

Channel 14

 

7

CH6_NO

15

CH14_NO

22

CH6_COM

30

CH14_COM

Channel 7

 

Channel 15

 

8

CH7_NO

43

CH15_NO

23

CH7_COM

44

CH15_COM

 

 

31 - 42

28

Image 30
Contents E2251-90002 Page Contents Contents Agilent Technologies Warranty Statement Documentation History Page Declaration of Conformity Page Page Module Installation Remove the Carrier Shields Determine M-Module Layout Set the Logical Address for each M-Module Guidelines for Determining Logical AddressesModule M3 Module M2 Module M1 Install M-Modules onto the Carrier Connect Field Wiring to Slot M4 & M5 M-Modules Reattach the Carrier Shields Install Carrier into C-Size VXI Mainframe Connect Field Wiring to Front M-Modules Installation / Operational Verification Removing M-Modules from the Carrier Module Field Wiring Information Assembling the M-Module Field Wiring ConnectorPage Place connector Connector screws on bottom shell E2259A Double-Wide Breadboard M-Module Wiring Information E2259A J102 Pinout Table E2261A Quad RS-232 Interface Wiring Information Serial Port 1 Serial Port Serial Port 3 Serial PortPage E2270A 16-Channel Form a Switch Wiring Information E2271A 4x4 Matrix Switch Wiring Information ROW2LOChannel Configuring the E2272A E2273A 8-Channel Form C Switch Wiring Information E2274A 4-Channel Power Relay Wiring Information E2290A 16-Bit Digital I/O Wiring Information UseE2291A 16-Channel Isolated Digital Output Wiring Information Page Index NumericsPosition in Carrier, 11 removing from Carrier, 20 slots