Step 5. Connect Field Wiring to Slot M4 & M5 M-Modules

If you installed M-Modules on either or both of the interior slots, the following instructions describe how to secure the field wiring. If you did not install an M-Module in slots M4 or M5, skip this page and go to Step 6.

1.Cover the M0 2-row connector strip with provided Pin Protector/Cover (p.n. 1252-8273). This pin protector protects the field wiring from chafing against the connector pins.

Note: Do not install the plastic cable hood (see page 21) on the field wiring connectors for the two internal M-Modules. Do not install the RFI dress panels on the two interior M-Modules. If an RFI dress panel is already installed on an M-Module, it must be removed. They are not needed and may interfere with installation of the top shield.

2.Holes are provided in the Carrier for cable ties to secure the field wiring cables.

3.Install cable clamp plates (p.n. E2251-01203) on top side of PC board. Secure the cable clamp (but do not overtighten). Both clamps are required to prevent damage to PC board.

4.To minimize RFI, install a dress panel (p.n. E2251-04301) over the front panel M0 slot position. Have the dress panel machined or punched to allow the cable to pass through. Machine the cut as small as possible; large enough for the cable and optional grommet (to prevent chafing) but tight to minimize RFI.

M3x20 Pan Head Screws p.n. 0515-1410

3. Secure field wiring to Carrier with wire ties through holes provided on Carrier.

 

2. Attach field wiring

 

to M-Modules

 

with 4-40 Pozidrive screws.

 

1. Place p.n. 1252-8273

 

Pin Protector over pins

 

to protect cable.

5. Machine Dress Panel

4. Install Cable Clamp.

Do not overtighten.

for ribbon cable or

 

punch Dress Panel and add grommet for round cable.

15

Image 17
Contents E2251-90002 Page Contents Contents Agilent Technologies Warranty Statement Documentation History Page Declaration of Conformity Page Page Module Installation Remove the Carrier Shields Determine M-Module Layout Guidelines for Determining Logical Addresses Set the Logical Address for each M-ModuleModule M3 Module M2 Module M1 Install M-Modules onto the Carrier Connect Field Wiring to Slot M4 & M5 M-Modules Reattach the Carrier Shields Install Carrier into C-Size VXI Mainframe Connect Field Wiring to Front M-Modules Installation / Operational Verification Removing M-Modules from the Carrier Assembling the M-Module Field Wiring Connector Module Field Wiring InformationPage Place connector Connector screws on bottom shell E2259A Double-Wide Breadboard M-Module Wiring Information E2259A J102 Pinout Table Serial Port 1 Serial Port Serial Port 3 Serial Port E2261A Quad RS-232 Interface Wiring InformationPage E2270A 16-Channel Form a Switch Wiring Information ROW2LO E2271A 4x4 Matrix Switch Wiring InformationChannel Configuring the E2272A E2273A 8-Channel Form C Switch Wiring Information E2274A 4-Channel Power Relay Wiring Information Use E2290A 16-Bit Digital I/O Wiring InformationE2291A 16-Channel Isolated Digital Output Wiring Information Page Numerics IndexPosition in Carrier, 11 removing from Carrier, 20 slots