Agilent Technologies E2251A user manual E2274A 4-Channel Power Relay Wiring Information

Page 35

E2274A 4-Channel Power Relay Wiring Information

MAXIMUM VOLTAGE/CURRENT: The maximum voltage that may be applied to any connector on the E2274A is 125 VDC, 141 VAC rms, or 200 VAC peak. These limits apply only if the product is installed in a humidity-con- trolled (<60% RH) environment where airborne contami- nants and transients are controlled, and there is NOT a relay connection made to power mains. If these condi- tions CANNOT be maintained, then the maximum volt- age is 60 VDC, 48 VAC rms or 68 VAC peak.

The maximum current (non-inductive) that may be applied to the E2274A is 5 ADC, 5 AACpeak per switch. Maximum power is 100W DC, 100VA AC, per switch; 300WDC, 300VA AC per M-module.

NC = Normally Closed

 

NO = Normally Open

No Connection: Pins 7, 19, 26

com = Common

CGND: Pins 14, 15, 30, 37, 44

Caution All three connector pins for each relay contact must be connected together in the field wiring for maximum current capability. For example, for Channel 1, pins 1, 16, and 31 (for NC contacts) must be wired together in the field wiring, likewise pins 3, 18, and 33 (for the NO contact) and pins 2, 17, and 32 (for channel common). You may need to run separate wires from each connector pin and unite them at the device under test or in your cable.

33

Image 35
Contents E2251-90002 Page Contents Contents Agilent Technologies Warranty Statement Documentation History Page Declaration of Conformity Page Page Module Installation Remove the Carrier Shields Determine M-Module Layout Guidelines for Determining Logical Addresses Set the Logical Address for each M-ModuleModule M3 Module M2 Module M1 Install M-Modules onto the Carrier Connect Field Wiring to Slot M4 & M5 M-Modules Reattach the Carrier Shields Install Carrier into C-Size VXI Mainframe Connect Field Wiring to Front M-Modules Installation / Operational Verification Removing M-Modules from the Carrier Assembling the M-Module Field Wiring Connector Module Field Wiring InformationPage Place connector Connector screws on bottom shell E2259A Double-Wide Breadboard M-Module Wiring Information E2259A J102 Pinout Table Serial Port 1 Serial Port Serial Port 3 Serial Port E2261A Quad RS-232 Interface Wiring InformationPage E2270A 16-Channel Form a Switch Wiring Information ROW2LO E2271A 4x4 Matrix Switch Wiring InformationChannel Configuring the E2272A E2273A 8-Channel Form C Switch Wiring Information E2274A 4-Channel Power Relay Wiring Information Use E2290A 16-Bit Digital I/O Wiring InformationE2291A 16-Channel Isolated Digital Output Wiring Information Page Numerics IndexPosition in Carrier, 11 removing from Carrier, 20 slots