Agilent Technologies E2251A user manual Removing M-Modules from the Carrier

Page 22

Removing M-Modules from the Carrier

WARNING To avoid possible hazardous electrical shock, remove power from the VXI Mainframe, all field wiring, and external devices connected to the VXI mainframe.

1.Remove all field wiring connectors from the front-mounted M-Modules of the E2251A. Refer to “Step 8. Connect Field Wiring to Front M-Modules” on page 18 for general guidelines.

2.Remove the Carrier from the VXI Mainframe. Open the extraction levers to release the Carrier from the mainframe. Refer to “Step 7. Install Carrier into C-Size VXI Mainframe” on page 17 for general guidelines.

3.Follow instructions in “Step 1. Remove the Carrier Shields” on page 10 of this manual to remove the top and bottom shields from the Carrier.

4.Refer to the figure below to remove M-Modules from the Carrier.

Remove these four screws

to remove M-Module from Carrier.

Do NOT remove these two screws from M-Module!

If moving M-Module to slot M4 or M5, remove

RFI Dress Panel from M-Module

Caution STATIC ELECTRICITY. Static electricity is a major cause of component failure. To prevent damage to the electrical components in the Carrier and the M-Modules, observe anti-static techniques when installing or removing an M-Module from the Carrier or installing or removing the Carrier from the VXI mainframe.

20

Image 22
Contents E2251-90002 Page Contents Contents Agilent Technologies Warranty Statement Documentation History Page Declaration of Conformity Page Page Module Installation Remove the Carrier Shields Determine M-Module Layout Set the Logical Address for each M-Module Guidelines for Determining Logical AddressesModule M3 Module M2 Module M1 Install M-Modules onto the Carrier Connect Field Wiring to Slot M4 & M5 M-Modules Reattach the Carrier Shields Install Carrier into C-Size VXI Mainframe Connect Field Wiring to Front M-Modules Installation / Operational Verification Removing M-Modules from the Carrier Module Field Wiring Information Assembling the M-Module Field Wiring ConnectorPage Place connector Connector screws on bottom shell E2259A Double-Wide Breadboard M-Module Wiring Information E2259A J102 Pinout Table E2261A Quad RS-232 Interface Wiring Information Serial Port 1 Serial Port Serial Port 3 Serial PortPage E2270A 16-Channel Form a Switch Wiring Information E2271A 4x4 Matrix Switch Wiring Information ROW2LOChannel Configuring the E2272A E2273A 8-Channel Form C Switch Wiring Information E2274A 4-Channel Power Relay Wiring Information E2290A 16-Bit Digital I/O Wiring Information UseE2291A 16-Channel Isolated Digital Output Wiring Information Page Index NumericsPosition in Carrier, 11 removing from Carrier, 20 slots