Figure 4. Sensor Orientation

FLOW

FLOW

The electrodes in the Rosemount 8705 sensor are properly orientated when the two measurement electrodes are in the 3 and 9 o’clock positions, as shown on the right of Figure 4.

The electrodes in the Rosemount 8711 are properly orientated when the top of the sensor is either vertical or horizontal, as shown in Figure 5. Avoid any mounting orientation that positions the top of the sensor at 45° from the vertical or horizontal position.

Figure 5. Rosemount 8711 Mounting Position

45° Electrode Plane

45° Electrode Plane

8

Image 8
Contents Quick Installation Guide Management Asia Pacific Private LimitedPRE-INSTALLATION Mechanical ConsiderationsDimensions are in inches millimeters 10912.02 305Installation Procedures Identify Options and ConfigurationsHardware Jumpers/Switches Environmental ConsiderationsHandling Electrical ConsiderationsMounting Upstream/Downstream PipingFlow Direction Sensor OrientationFlow Installation Flanged SensorsGaskets Flange BoltsPound-feet Size ClassCode Newton Code Meter PTFE/ETFE liner PN10 SizeNewton Polyurethane, Linatex, Adiprene and Neoprene Liners Size Nominal Sensor Size Wafer SensorsAlignment Rosemount Alignment Spacer Table Line Size Dash No Pound-feet Newton-meter Sanitary SensorsAlignment and Bolting Grounding Process Grounding Options Type of PipeLining StrapsGrounding with Grounding Electrode Wiring Power OutputsCoil Drive and Electrode Cables Power Outputs Conduit Ports and ConnectionsTransmitter to Sensor Wiring Remote Mount Cable Requirements and PreparationWiring the Transmitter to the Sensor Transmitter Terminal Sensor Terminal Wire Gauge Wire ColorDescription Part NumberRemote Mount Wiring Diagrams Connect the 4-20 mA Analog Signal Internal Power SourceExternal Power Source Cabling considerationsPowering the Transmitter Supply Wire RequirementsInstallation Category DisconnectsFuse 8712E Power SupplyAC Ground or DC Ground Transmitter Power Cable Basic Setup Basic Setup Local Operator InterfaceFunction Fast Keys Process Variables PV ReviewProduct Certifications Approved Manufacturing LocationsEuropean Directive Information International CertificatesEuropean Certifications N1 Atex Type n Special Conditions for Safe UseInternational Certifications N7 IECEx Type nSensor Approval Information Maximum Process Meter Size Inches Maximum Ambient TemperatureRosemount 8705 and 8711 Sensors EC Declaration of Conformity Schedule