Teledyne MX300 manual Sensor Installation or Replacement, Installing the R17MED Sensor

Page 16

Operation

MX300

 

 

 

2.1.1 Sensor Installation or Replacement

Note: The R17MED oxygen sensor must be installed before the oxygen analyzer/monitor can be operated

Remove the new sensor from its protective bag. Inspect the sensor for damage or electrolyte leakage. If the sensor is damaged, obtain a replacement. Do not use the defective sensor as it may damage the unit.

WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT LET IT COME IN CONTACT WITH SKIN. IF IT DOES, FLUSH AFFECTED AREA WITH WATER. DO NOT ATTEMPT TO OPEN OR REPAIR THE SENSOR.

WARNING: THE SENSOR ALSO CONTAINS LEAD. LEAKING OR EXHAUSTED SENSORS SHOULD BE HANDLED AND DISPOSED OF IN ACCORDANCE WITH LOCAL REGULATIONS. CONTACT TELEDYNE FOR THE MATERIAL SAFETY DATA SHEET

2.Plug one end of the coiled cable into the jack receptacle on the back end of the R17MED sensor and secure in place with the capture nut located at the base of the connector. See Figure 2-1.

Figure 2-1: Installing the R17MED Sensor

3.Plug the other end of the coiled cable into the receptacle located on the right side of the unit and secure it in place using the capture nut. See Figure 2-2.

Teledyne Analytical Instruments

16

Image 16
Contents MX300 MX300 How To Use This Manual About This ManualSafety Messages Table of Contents Appendix List of Figures MX300 Front ViewList of Tables Introduction MX300 Front ViewApplicable Standards Features OptionsApplications Theory of OperationSignal Processing SensorIntroduction Setup OperationSensor Installation or Replacement Installing the R17MED SensorSensor Cable Connection to Monitor V-Mount Adapter Installation MountingBattery Installation Installing BatteriesCalibration Calibration SPortable Oxygen Monitor Operation Alarms Setting the AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Procedure UseGas Sampling HumidityTemperature PressureDiscrepancy in Readings Anesthetic GasesOxygen Reading Error in a Mixture of Anesthetic Gas Cleaning InstrumentDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Overall Maintenance General Service InformationBattery Maintenance Sensor Maintenance CalibrationAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Error Codes WatchdogOther Problems with the Instrument Return Authorization for Service Specifications AppendixSpare Parts List Optional AccessoriesOrders should be sent to Blank Index Index