Teledyne MX300 manual √ Sensor

Page 40

Service Manual

MX300

 

 

 

Symptom

Why

What To Do

The sensor does not

Water is condensing

A) Remove the sensor from tee

react to changes in

on the sensing

adapter and unscrew the

oxygen concentration,

surface.

plastic flow diverter. Using

or the readings are

Electrical interference

absorbent tissue or cotton

unstable and drifting.

swab, gently wipe off sensing

is disrupting the

 

surface inside threaded

 

electronics

 

portion of sensor assembly.

 

 

 

 

B) Relocate unit away from

 

 

sources of electrical noise

 

 

such as cauterizing equipment

 

 

and two-way radios.

The display is flashing

The unit has detected

A) Check sensor cable

√ SENSOR

a fault in the signal

connections and make sure

from the sensor.

they are completely inserted

 

 

Sensor has expired.

into the mating connector

 

and the capture nut is firmly

 

 

 

The sensor has been

in place.

 

exposed to a gas

B) Expose the sensor in 100%

 

containing little or no

 

O2 and check calibration.

 

oxygen.

 

C) The sensor output has fallen

 

 

 

 

to a level where it is no longer

 

 

usable. Replace sensor.

The oxygen reading

Like all O2 sensors,

A) During calibration, make

fluctuates or appears to

the R17MED detects

sure there is no restrictions

be incorrect.

the changes in the

on exhaust side of sensor. If

 

partial pressure of

the reading changes with

 

O2.

flow, the sensor is

 

 

pressurized or there may be

 

 

a leak in the system.

 

 

B) If a high degree of accuracy

 

 

is desired, or the

 

 

concentration of O2 is in

 

 

excess of 40%, calibration

 

 

with 100% is recommended.

 

 

C) If humidified gas is used to

 

 

ventilate the patient, water

 

 

vapor actually dilutes the

 

 

gas. See Appendix:

 

 

Humidity, Temperature.

 

 

D) If a blender is used, check

 

 

its calibration. See

 

 

Appendix: Discrepancy in

 

 

Readings.

Teledyne Analytical Instruments

40

Image 40
Contents MX300 MX300 How To Use This Manual About This ManualSafety Messages Table of Contents Appendix List of Figures MX300 Front ViewList of Tables Introduction MX300 Front ViewApplicable Standards Features OptionsApplications Theory of OperationSignal Processing SensorIntroduction Setup OperationSensor Installation or Replacement Installing the R17MED SensorSensor Cable Connection to Monitor V-Mount Adapter Installation MountingBattery Installation Installing BatteriesCalibration Calibration SPortable Oxygen Monitor Operation Alarms Setting the AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Procedure UseGas Sampling HumidityTemperature PressureDiscrepancy in Readings Anesthetic GasesOxygen Reading Error in a Mixture of Anesthetic Gas Cleaning InstrumentDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Overall Maintenance General Service InformationBattery Maintenance Sensor Maintenance CalibrationAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Error Codes WatchdogOther Problems with the Instrument Return Authorization for Service Specifications AppendixSpare Parts List Optional AccessoriesOrders should be sent to Blank Index Index