Teledyne MX300 manual Do’s and Don’ts

Page 31

Portable Oxygen Monitor

Operation

 

 

 

manufacturer’s instructions). Due to the varying conditions imposed on the materials during sterilization it is not possible to determine the exact number of times the sterilization processes can be carried out. Therefore, Teledyne recommends that the operators carefully examine the diverter and tee adapter after sterilization and prior to use to verify that the item is fit for use. The operator should verify that there are no cracks and tears and the item does not show any indication of material changes or physical damage that may compromise its effective use. Both diverter and tee adapter should be free of any chemical residue attributable to the sterilization process.

Because of the variability of the cleaning, disinfecting and sterilizing processes, Teledyne cannot provide specific sterilization instructions nor can the sterility of the item be ensured. Therefore, we highly recommend referring to the manufacturer’s instructions on the details of method.

2.4 Do’s and Don’ts

– DO –

Read all of the directions before using for the first time.

Calibrate every 8 hours or before every use.

Visually inspect the sensor for leakage before each use.

Calibrate using 100% oxygen and check in air.

Check the HI and LO alarm settings prior to each use.

Test batteries regularly and replace when battery indicator shows low battery (no bars remaining)

Make sure keys are locked by using the LOCK/UNLOCK key feature.

Keep the unit, sensor and connections dry, or on the dry side of the breathing circuit.

Recalibrate after replacing the batteries.

Teledyne Analytical Instruments

31

Image 31
Contents MX300 MX300 About This Manual How To Use This ManualSafety Messages Table of Contents Appendix MX300 Front View List of FiguresList of Tables MX300 Front View IntroductionApplicable Standards Options FeaturesTheory of Operation ApplicationsSensor Signal ProcessingIntroduction Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Monitor Mounting V-Mount Adapter InstallationInstalling Batteries Battery InstallationCalibration S CalibrationPortable Oxygen Monitor Operation Setting the Alarms AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Use ProcedureHumidity Gas SamplingPressure Temperature Anesthetic Gases Discrepancy in ReadingsOxygen Reading Error in a Mixture of Anesthetic Gas Instrument CleaningDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Overall Maintenance General Service InformationBattery Maintenance Calibration Sensor MaintenanceAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service Appendix SpecificationsOptional Accessories Spare Parts ListOrders should be sent to Blank Index Index