Teledyne MX300 manual Sensor Maintenance, Calibration

Page 36

Service Manual

MX300

 

 

 

The MX300 instrument incorporates a battery test feature that provides the approximate amount of useful life remaining on the set of batteries. Excessive alarm activation will wear down the battery faster than usual.

The minimum detectable change in battery voltage corresponds to an increment of about 50 hours, meaning that the battery voltage reading may not change for several hours at a time.

If the monitor is not used for a period of 30 days or more, the batteries should be removed prior to storage.

3.4 Sensor Maintenance

DO: check the sensor for damage or leaks before use.

DO: recalibrate after replacing the sensor.

DON’T: immerse the R17MED sensor in liquid. DON’T autoclave the R17MED sensor. DON’T: open or try to repair the sensor.

Before every use, the sensor, cable and connections should be checked. Check the sensor for leaks and condensation. Check the cable for splitting or cracked insulation. Make sure the connections are tight and dry.

In the event that the sensor has been damaged, consult the Material Safety Data Sheet in the Appendix for handling guidelines.

3.5 Calibration

Incorrect readings can often traced to improper calibration. The MX300 should be calibrated before each use and every 8 hours to maintain accuracy. It must be calibrated whenever new batteries are installed. Calibration using methods other than described below can lead to improper operation and are discouraged.

Whenever new batteries are installed or removed for any reason, the oxygen monitor defaults to the calibration mode with the LCD display flashing 00.0. Only the CAL and ON/OFF keys are functional at this point. After calibration is completed, the alarm set points will need to be reset.

To calibrate the instrument refer back to Section 2.1.3.

Teledyne Analytical Instruments

36

Image 36
Contents MX300 MX300 How To Use This Manual About This ManualSafety Messages Table of Contents Appendix List of Figures MX300 Front ViewList of Tables Introduction MX300 Front ViewApplicable Standards Features OptionsApplications Theory of OperationSignal Processing SensorIntroduction Setup OperationSensor Installation or Replacement Installing the R17MED SensorSensor Cable Connection to Monitor V-Mount Adapter Installation MountingBattery Installation Installing BatteriesCalibration Calibration SPortable Oxygen Monitor Operation Alarms Setting the AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Procedure UseGas Sampling HumidityTemperature PressureDiscrepancy in Readings Anesthetic GasesOxygen Reading Error in a Mixture of Anesthetic Gas Cleaning InstrumentDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments General Service Information Overall MaintenanceBattery Maintenance Sensor Maintenance CalibrationAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Error Codes WatchdogOther Problems with the Instrument Return Authorization for Service Specifications AppendixSpare Parts List Optional AccessoriesOrders should be sent to Blank Index Index