Teledyne MX300 manual Output 0-1 VDC or RS232

Page 23

Portable Oxygen Monitor

Operation

 

 

 

you to select a LO alarm value less than 18%.

Note: The MX300 Portable Oxygen Monitor is designed to prevent crossing of HI/LO alarm settings. If you attempt to set the LO alarm higher than the HI alarm, it will push the HI alarm setting up as you continue to rise the LO alarm set point. This also applies when attempting to set the HI alarm lower than the LO alarm.

3.To test the alarms, unlock the controls by pressing the LOCK/UNLOCK key and then press ALARM TEST key. The Lo alarm followed by the HI alarm will blink and the audible visual alarm will are activated for about 1 second.

4.To silence the alarm buzzer for 115 seconds press the ALARM SILENCE key (flashing red indicator light).

5.To disable the alarms set the high alarm set point above 100%.

Note: When the alarm is in the OFF condition (set point above 100%) the ALARM OFF status message blinks slowly on the LCD below the oxygen readout.

2.1.5 Output 0-1 VDC or RS232

The MX300 provides signal outputs for use with recorders and computers. The instruments are supplied standard with a 0-1 VDC output. An optional 0-1 VDC Interface Cable (P/N B-75554) is available from Teledyne for this purpose.

To connect the analyzer to an analog recording device:

1.Insert one end of the interface cable into the output port on the side of the instrument. See Figure 2-9.

2.Insert the other end into the analog recorder device. Make sure the device is equipped to handle a 0-1 VDC signal.

When properly calibrated, the output signal generated by the analyzer is linear and proportional to the oxygen concentration.

If you requested Option-B (RS 232 digital output) at the time of purchase, a digital RS 232 signal is output from the output port shown in Figure 2-8. Use the optional RS 232 Interface Cable (P/N B-75555) available from Teledyne for connection to a standard RS 232 port on a computer or other suitably equipped digital device.

Teledyne Analytical Instruments

23

Image 23
Contents MX300 MX300 About This Manual How To Use This ManualSafety Messages Table of Contents Appendix MX300 Front View List of FiguresList of Tables MX300 Front View IntroductionApplicable Standards Options FeaturesTheory of Operation ApplicationsSensor Signal ProcessingIntroduction Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Monitor Mounting V-Mount Adapter InstallationInstalling Batteries Battery Installation Calibration S CalibrationPortable Oxygen Monitor Operation Setting the Alarms AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Use ProcedureHumidity Gas SamplingPressure TemperatureAnesthetic Gases Discrepancy in ReadingsOxygen Reading Error in a Mixture of Anesthetic Gas Instrument CleaningDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Battery Maintenance General Service InformationOverall Maintenance Calibration Sensor MaintenanceAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service Appendix SpecificationsOptional Accessories Spare Parts ListOrders should be sent to Blank Index Index