Portable Oxygen Monitor | Service Manual | |
|
|
|
Symptom | Why | What To Do |
The unit has stopped | The MX300 | A) Disconnect the batteries and |
working and the LCD is | instruments are | inspect the contacts for |
displaying | equipped with an | corrosion. Reconnect the |
alphanumeric figures. | electronic "watch dog," | batteries. If the unit functions |
| which monitors the | properly, calibrate the unit |
| circuitry within the unit | and reset the alarm values. |
| for potential faults and | B) Try a new set of batteries. |
| renders the unit | |
|
| |
| inoperable until the | C) Increase the distance |
| condition is corrected. | between the unit and any |
| (See Watch dog | source of radio frequency |
| section below) | interference. The sensor |
| Several conditions can | cable is a prime source of |
| pickup as it can act like an | |
| activate the "watch | |
| antenna. Relocate the | |
| dog." Dropping the | |
| sensor cable and if possible | |
| unit, poor battery | |
| change its coiled length to | |
| connections, and radio | |
| ||
| frequency interference | |
| Placing the cable in a | |
| are the most common | |
| different position may also | |
| causes. See the watch | |
| help. | |
| dog section for | |
|
| |
| additional information. |
|
|
|
|
Alarm sounds/flashes | A) Readings are | A) Adjust high and low alarm |
continuously. | outside alarm | setting to be above and |
| limits. | below O2 value being |
|
| displayed. |
No display. | A) Batteries expired. | A) Check/replace batteries. |
| B) Bad battery | B) Check battery connections. |
| connection. | C) Calibrate. |
|
| |
Keys inoperable/cannot | The LOCK/UNLOCK | A) Press LOCK/UNLOCK key |
turn unit off | key is activated which | once. LCD will flash |
| is preventing key | indicating keys are active. |
| operation |
|
Cannot adjust | Critical settings require | A) If display is not flashing |
calibration or alarm | two keys be pressed in | press lock key once to |
settings | a specific order. | activate keys. Press desired |
| LOCK/UNLOCK key is | function followed by the Up |
| and Down key. | |
| active | |
|
| |
|
|
|
Note: In the event that none of these procedures produce desired results, remove the batteries and return the unit to Teledyne for repair.