Teledyne MX300 manual Portable Oxygen Monitor

Page 41

Portable Oxygen Monitor

Service Manual

 

 

 

Symptom

Why

What To Do

The unit has stopped

The MX300

A) Disconnect the batteries and

working and the LCD is

instruments are

inspect the contacts for

displaying

equipped with an

corrosion. Reconnect the

alphanumeric figures.

electronic "watch dog,"

batteries. If the unit functions

 

which monitors the

properly, calibrate the unit

 

circuitry within the unit

and reset the alarm values.

 

for potential faults and

B) Try a new set of batteries.

 

renders the unit

 

 

 

inoperable until the

C) Increase the distance

 

condition is corrected.

between the unit and any

 

(See Watch dog

source of radio frequency

 

section below)

interference. The sensor

 

Several conditions can

cable is a prime source of

 

pickup as it can act like an

 

activate the "watch

 

antenna. Relocate the

 

dog." Dropping the

 

sensor cable and if possible

 

unit, poor battery

 

change its coiled length to

 

connections, and radio

 

“de-tune” its antenna effect.

 

frequency interference

 

Placing the cable in a

 

are the most common

 

different position may also

 

causes. See the watch

 

help.

 

dog section for

 

 

 

additional information.

 

 

 

 

Alarm sounds/flashes

A) Readings are

A) Adjust high and low alarm

continuously.

outside alarm

setting to be above and

 

limits.

below O2 value being

 

 

displayed.

No display.

A) Batteries expired.

A) Check/replace batteries.

 

B) Bad battery

B) Check battery connections.

 

connection.

C) Calibrate.

 

 

Keys inoperable/cannot

The LOCK/UNLOCK

A) Press LOCK/UNLOCK key

turn unit off

key is activated which

once. LCD will flash

 

is preventing key

indicating keys are active.

 

operation

 

Cannot adjust

Critical settings require

A) If display is not flashing

calibration or alarm

two keys be pressed in

press lock key once to

settings

a specific order.

activate keys. Press desired

 

LOCK/UNLOCK key is

function followed by the Up

 

and Down key.

 

active

 

 

 

 

 

Note: In the event that none of these procedures produce desired results, remove the batteries and return the unit to Teledyne for repair.

Teledyne Analytical Instruments

41

Image 41
Contents MX300 MX300 About This Manual How To Use This ManualSafety Messages Table of Contents Appendix MX300 Front View List of FiguresList of Tables MX300 Front View IntroductionApplicable Standards Options FeaturesTheory of Operation ApplicationsSensor Signal ProcessingIntroduction Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Monitor Mounting V-Mount Adapter InstallationInstalling Batteries Battery InstallationCalibration S CalibrationPortable Oxygen Monitor Operation Setting the Alarms AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Use ProcedureHumidity Gas SamplingPressure TemperatureAnesthetic Gases Discrepancy in ReadingsOxygen Reading Error in a Mixture of Anesthetic Gas Instrument CleaningDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Battery Maintenance General Service InformationOverall Maintenance Calibration Sensor MaintenanceAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service Appendix SpecificationsOptional Accessories Spare Parts ListOrders should be sent to Blank Index Index