Portable Oxygen Monitor | Service Manual | |
|
|
|
Note: The MSDS on this material is located in the Appendix. It is also available upon request through the Teledyne Environmental, Health and Safety Coordinator. Contact at
(626)
3.8Troubleshooting
The MX300 oxygen monitor provides a variety of
Table 3-1 Troubleshooting
Symptom | Why | What To Do |
New sensor responds | If the sensor is new | A) Wait |
slow or drifts. | and was just | recalibrate. |
| removed from its |
|
| sealed bag it may |
|
| need to run for |
|
| several hours. |
|
Sensor will not read | Calibration in 100% | A) Recalibrate using dry gas |
below 22 % after | was invalid or the | making sure the reading |
calibration in 100% O2. | room air is | stabilizes before making |
| contaminated with | any adjustments. |
| excess oxygen. | B) Make sure that at least 6" |
|
| |
|
| (30 cm.) of tubing is |
|
| attached to the exhaust |
|
| side of the tee adapter to |
|
| prevent back filling. O2 flow |
|
| rate should not exceed 5 |
|
| l/min. |
|
| C) Oxygen concentration at the |
|
| sensor is significantly higher |
|
| than 21%. Take the |
|
| instrument to a well- |
|
| ventilated area and check |
|
| the reading again. |
|
| D) Try calibrating with a known |
|
| good sensor; if this fails, see |
|
| symptom "Reading drifts |
|
| over |
Teledyne Analytical Instruments | 39 |