Teledyne MX300 manual Use, Procedure

Page 25

Portable Oxygen Monitor

Operation

 

 

 

2.2 Use

2.2.1 Procedure

Note: Prior to use, always test the batteries and alarms (if applicable). Also check calibration, the sensor for leaks and damage, and the alarm settings.

The AX/MX300 instruments can be used to measure a gas mixture for oxygen in two basic modes:

In the inhalation side of breathing circuit ahead of antibacterial filters, humidifiers and medicating devices.

In confined volumes such as incubators or tents.

When monitoring for oxygen in breathing circuits, the flow diverter must be used. The diverter should be screwed onto the threaded front end of the R17MED sensor. A tee adapter (plastic, P/N A268, or metal, P/N A283) should be placed into the circuit, and the above sensor assembly plugged into the tee adapter. See Figure 2-1.

CAUTION: CHECK THE BREATHING CIRCUIT FOR LEAKS. BE CERTAIN THAT THE CIRCUIT DOWNSTREAM OF THE SENSOR DOES NOT PRODUCE ANY BACKPRESSURE OR RESTRICTION TO FLOW. ERRORS IN READINGS WILL RESULT IF THIS IS NOT FOLLOWED.

THE OXYGEN SENSOR IS A NON-STERILE DEVICE AND SHOULD BE USED IN CONJUCTION WITH AN ANTIBACTERIAL FILTER. ALWAYS INSTALL THE SENSOR ON THE INSPIRED LINE AHEAD OF FILTERS, HUMIDIFIERS AND MEDICATING DEVICES. NEVER EXPOSE THE SENSOR TO PATIENT’S EXHALED BREATH OR SECRETIONS UNLESS YOU INTEND TO DISPOSE OF THE SENSOR AND FLOW ADAPTER AFTER USE.

When monitoring for oxygen in confined volumes such as incubators, hoods, etc., the flow diverter must be removed from the

Teledyne Analytical Instruments

25

Image 25
Contents MX300 MX300 About This Manual How To Use This ManualSafety Messages Table of Contents Appendix MX300 Front View List of FiguresList of Tables MX300 Front View IntroductionApplicable Standards Options FeaturesTheory of Operation ApplicationsSensor Signal ProcessingIntroduction Operation SetupInstalling the R17MED Sensor Sensor Installation or ReplacementSensor Cable Connection to Monitor Mounting V-Mount Adapter InstallationInstalling Batteries Battery InstallationCalibration S CalibrationPortable Oxygen Monitor Operation Setting the Alarms AlarmsOutput 0-1 VDC or RS232 0-1 VDC or RS 232 Digital Output Port Use ProcedureHumidity Gas SamplingPressure TemperatureAnesthetic Gases Discrepancy in ReadingsOxygen Reading Error in a Mixture of Anesthetic Gas Instrument CleaningDo’s and Don’ts Operation Portable Oxygen Monitor Operation Blank Teledyne Analytical Instruments Overall Maintenance General Service InformationBattery Maintenance Calibration Sensor MaintenanceAlarms Pressure Troubleshooting Troubleshooting√ Sensor Portable Oxygen Monitor Watchdog Error CodesOther Problems with the Instrument Return Authorization for Service Appendix SpecificationsOptional Accessories Spare Parts ListOrders should be sent to Blank Index Index