Portable Oxygen Monitor | Operation | |
|
|
|
Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas
Gas or Vapor Level
(Balance: Mixture of 30% O2 / 70% N2O, except where noted)
Gas or Vapor | Test Level | Oxygen Reading |
|
| Error |
Helium | 50%, balance O2 | 0% |
Nitrous Oxide | 80%, balance O2 | 0% |
Carbon Dioxide | 10%, balance O2 | 0% |
Halothane | 4% | < 1.5% O2 * |
Enflurane | 5% | < 1.5% O2 * |
Isoflurane | 5% | < 1.5% O2 * |
Sevoflurane | 5% | < 1.5% O2 * |
Desflurane | 15% | < 1.5% O2 * |
*Errors are approximate and may vary based on exposure times and concentrations.
These performances meet or exceed the requirements of ISO 7767: 1997 (E).
CAUTION: THE MX300 SHOULD NOT BE USED IN THE PRESENCE OF FLAMMABLE ANESTHETICS SUCH AS DIETHYL ETHER OR CYCLOPROPANE.
CAUTION: THE MX300, OXYGEN SENSOR AND ASSOCIATED HARDWARE ARE
2.3.5.2CARE AFTER USE IN NITROUS OXIDE
CAUTION: THE R17MED SENSOR SHOULD NOT BE LEFT IN NITROUS OXIDE MIXTURES ANY LONGER THAN ABSOLUTELY NECESSARY.
Teledyne Analytical Instruments | 29 |