Teledyne 356WA Trace Oxygen Analyzer Installation, Location, Electrical Connections

Page 11

Trace Oxygen Analyzer

Installation 3

 

 

 

 

Installation

3.1 Location

With proper shielding of the leads, the analyzer and the readout device can be separated by as much as 1,000 feet. However, they should be placed as close together as possible. For the most convenient operation, the read- out recorder or meter should be within view of the controls, particularly when the unit is being calibrated. Figure 2-1 depicts a typical system layout. Other location considerations:

1)The analyzer should be sheltered from the elements.

2)Ambient temperature must be within 30 to 120 °F.

3)The unit should not be subject to excessive shock or vibration.

4)It should be as close as possible to the sample point.

5)There must be access to the back and side of the unit for connection or maintenance of sample lines and power.

NOTE: Since the level of the electrolyte in the measuring cell is critical and the water level control system for the humidifier is gravity sensitive,

THE ANALYZER MUST BE MOUNTED SO THAT THE BOTTOM OF THE CASE IS LEVEL.

3.2 Electrical Connections

A diagram of the necessary electrical connections is shown in Figure 2-2.

Note: See the Interconnection Diagram (drawing A-37526) included in the back of this manual, as well as any Addenda that may be included with this manual for information specific to your instrument.

Teledyne Analytical Instruments

3-1

Image 11
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 356WAOperational Theory Trace Oxygen Analyzer Table of Contents IntroductionInstallation Maintenance & TroubleshootingAppendix Method of Operation Trace Oxygen Analyzer IntroductionRecorder /Meter Readout Required EquipmentSample Conditioning Humidifier SensorFlow System Flow System Schematic Operational Theory Electrical Connections Trace Oxygen Analyzer InstallationLocation Typical System Layout Sample ConnectionsGND Gas Connections to Back of Analyzer Cell Packaging Trace Oxygen Analyzer OperationsFilling the Reservoir Detector CellTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Humidity Control Throttle ValveCalibrator Assembly Calibration PowerWarm-Up and Stabilization Operations Reservoir Flowmeter and HumidifierCell Electrolyte Level Lead Electrode Electrolyte ReplacementCell Cell Assembly Reservoir and Humidifier Column Screen AssemblyRemoving the Humidifier Column To ReinstallCell Leak Leak Detection ProcedureLeak Detection Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do What To Do Symptom CauseTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications QTY Description Spare Parts ListDrawing List Trace Oxygen Analyzer AppendixCalibration Considerations Calibration DataDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix From 288-308K Ambient Temperature K Vapor Pressure of Water mm HgVapor Pressure of Water