Trace Oxygen Analyzer | Appendix |
|
|
|
|
When span gases are used or when sample gas is being analyzed the flow can vary
If positioning the float in the center of the target is in error by plus or minus
There are instances when it will be necessary to reset or check the exit flow of sample gas and set or reset the position of the target on the flow tube:
•When a different background gas is being analyzed.
•When the target has been accidently repositioned on the flow tube.
The vent flow should be measured either with a volume displacement flow device (e.g. a
To determine the corrected vent flow rate it will be necessary to know the ambient temperature (in °K), the ambient pressure (in mm Hg) and the vapor pressure of water at ambient temperature.
Ambient temperature can be measured in °Centigrade or °F and con- verted to °K.
Degrees C = 5 (°F -32)
9
Degrees K = °C + 273
Ambient pressure can be measured with an accurate barometer.
P(mm Hg) = P(in Hg) x 760 30.00
*If the actual oxygen concentration were, for example, 8.0 ppm, the resultant reading would be 7.9 (float high) or 8.1 (float low).