4 Operations

Model 356WA

 

 

 

 

To calibrate the analyzer:

1.Shut the sample gas off, disconnect the sample in line and connect the the span gas to the sample inlet. Allow the span gas to flow through the system. Adjust the span gas flow rate until the flowmeter float is centered in the flowmeter reference indicator. Note the reading of the external recorder or meter.

2.Turn the range selector switch to the range which encompases the oxygen concentration of the span gas. The meter or recorder should move upscale and indicate the exact concentration of oxygen in the span gas.

3.If the external recorder/meter does not indicate the proper amount of oxygen, adjust the span dial until it does.

4.After the recorder or meter has stabilized on the proper readout, turn off the span gas flow, disconnect the span gas, reconnect the sample gas and establish the proper flow.

4-8

Teledyne Analytical Instruments

Image 22
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 356WA Copyright 1999 Teledyne Analytical InstrumentsMaintenance & Troubleshooting Trace Oxygen Analyzer Table of Contents IntroductionInstallation Operational TheoryAppendix Trace Oxygen Analyzer Introduction Method of OperationSample Conditioning Required EquipmentRecorder /Meter Readout Sensor HumidifierFlow System Flow System Schematic Operational Theory Location Trace Oxygen Analyzer InstallationElectrical Connections Sample Connections Typical System LayoutGND Gas Connections to Back of Analyzer Detector Cell Trace Oxygen Analyzer OperationsFilling the Reservoir Cell PackagingTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Throttle Valve Humidity ControlCalibrator Assembly Warm-Up and Stabilization PowerCalibration Operations Cell Electrolyte Level Flowmeter and HumidifierReservoir Cell Electrolyte ReplacementLead Electrode Cell Assembly Screen Assembly Reservoir and Humidifier ColumnTo Reinstall Removing the Humidifier ColumnLeak Detection Leak Detection ProcedureCell Leak Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do Symptom Cause What To DoTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications Spare Parts List QTY DescriptionTrace Oxygen Analyzer Appendix Drawing ListCalibration Data Calibration ConsiderationsDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix Vapor Pressure of Water Ambient Temperature K Vapor Pressure of Water mm HgFrom 288-308K