Teledyne 356WA instruction manual Teledyne Analytical Instruments

Page 30

5 Maintenance & Troubleshooting

Model 356WA

 

 

 

 

 

 

 

 

 

 

Symptom

Cause

What To Do

No analyzer response to

a)

Poor electrical connection,

oxygen.

 

or F1 fuse (a standard 2

 

 

amp Slo-Blo fuse) has

 

 

blown.

 

b)

Cell electrolyte level too

 

 

low.

 

c)

Dirty cell.

 

d)

Short between cell cathode

 

 

and anode (the screen to

 

 

lead [the base material] is

 

 

shorted.)

 

e)

If d) above corrects the

 

 

problem, the cell has

 

 

probably been poisoned,

 

 

probably by fluid flowing into

 

 

the cell humidifier column.

 

f)

Cell has been poisoned by a

 

 

component in the sample.

a)Verify proper connection from the cell through the control unit to the external recorder or meter; check for blown fuse. Replace as necessary.

b)Inspect electrolyte level and add as necessary. See Section 4.2.3, item 4.

c)Remove cell and clean thoroughly. See Section 4.2.3.

d)Correct short. Refill with fresh electrolyte as needed.

e)Check for excessive flow- rate.

Check for excessive foam- ing in the humidifier column. Drain and clean reservoir as necessary. Refer to Section 5.8.

f)If there has been no change in the normal sample composition, a scrubber may be required to remove the offending component.

Cell lacks sensitivity.

a)

Cell electrolyte level too low

 

 

due to misadjusted or faulty

 

 

humidity control.

 

b)

Cell electrolyte too high,

 

 

due to misadjusted or faulty

 

 

humidity control.

 

c)

Faulty humidity control.

a)Add electrolyte as neces- sary. Adjust humidity control. See Section 4.5.

b)Drain electrolyte and refill. Adjust humidity control. See Section 4.5.

c)• Shut off main power.

Disconnect orange wire from TS1 terminal 5, place a voltmeter (set to AC current, range 0–100 mA) in series with TS1-5 and the disconnected orange wire.

CAUTION: High voltage AC present.

5-8

Teledyne Analytical Instruments

Image 30
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 356WA Copyright 1999 Teledyne Analytical InstrumentsMaintenance & Troubleshooting Trace Oxygen Analyzer Table of Contents IntroductionInstallation Operational TheoryAppendix Trace Oxygen Analyzer Introduction Method of OperationRequired Equipment Sample ConditioningRecorder /Meter Readout Sensor HumidifierFlow System Flow System Schematic Operational Theory Trace Oxygen Analyzer Installation LocationElectrical Connections Sample Connections Typical System LayoutGND Gas Connections to Back of Analyzer Detector Cell Trace Oxygen Analyzer OperationsFilling the Reservoir Cell PackagingTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Throttle Valve Humidity ControlCalibrator Assembly Power Warm-Up and StabilizationCalibration Operations Flowmeter and Humidifier Cell Electrolyte LevelReservoir Electrolyte Replacement CellLead Electrode Cell Assembly Screen Assembly Reservoir and Humidifier ColumnTo Reinstall Removing the Humidifier ColumnLeak Detection Procedure Leak DetectionCell Leak Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do Symptom Cause What To DoTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications Spare Parts List QTY DescriptionTrace Oxygen Analyzer Appendix Drawing ListCalibration Data Calibration ConsiderationsDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix Ambient Temperature K Vapor Pressure of Water mm Hg Vapor Pressure of WaterFrom 288-308K