Teledyne 356WA instruction manual Symptom Cause, What To Do

Page 32

5 Maintenance & Troubleshooting

Model 356WA

 

 

 

 

Symptom

Cause

What To Do

Cell lacks sensitivity (contin-

Faulty humidity control (contin-

ued.)

ued.)

 

To replace humidifier control:

Shut off all power and remove the AC power cord from the power source.

Refer to wiring diagram D- 18633 (upper right-hand corner of the control unit.) Remove the wiring and controls. Replace R2 with appropriate part, P/N P31.

To replace humidifier heater element:

Shut off all power and remove the AC power cord from the power source.

Refer to wiring diagram C- 18593 (left-hand side, bottom of the cell com- partment.)

Pull out the block assem- bly from the bottom of the humidifier column; pull out the heater element, and unsolder the wires.

Replace with new heater element (P/N R139.) Solder wires, turn the terminals down into the holder, and replace the holder.

d)Cell compartment not at proper temperature (120 °F).

e)Faulty triac (Q6) in heater control or faulty triac driver (A4).

d)Verify that the power switch is ON.

Check condition of 2A fuse.

e)• Remove the cell compart- ment cover. Refer to Interconnection diagram A-18594 (upper left-hand corner.)

Connect voltmeter (set to high AC) to terminals 1 and 2 of terminal strip TS4 (Schematic diagram D-18632, lower right-hand corner; Wiring diagram C- 18593, upper left-hand corner.)

5-10

Teledyne Analytical Instruments

Image 32
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 356WA Copyright 1999 Teledyne Analytical InstrumentsTrace Oxygen Analyzer Table of Contents Introduction InstallationMaintenance & Troubleshooting Operational TheoryAppendix Trace Oxygen Analyzer Introduction Method of OperationRecorder /Meter Readout Required EquipmentSample Conditioning Sensor HumidifierFlow System Flow System Schematic Operational Theory Electrical Connections Trace Oxygen Analyzer InstallationLocation Sample Connections Typical System LayoutGND Gas Connections to Back of Analyzer Trace Oxygen Analyzer Operations Filling the ReservoirDetector Cell Cell PackagingTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Throttle Valve Humidity ControlCalibrator Assembly Calibration PowerWarm-Up and Stabilization Operations Reservoir Flowmeter and HumidifierCell Electrolyte Level Lead Electrode Electrolyte ReplacementCell Cell Assembly Screen Assembly Reservoir and Humidifier ColumnTo Reinstall Removing the Humidifier ColumnCell Leak Leak Detection ProcedureLeak Detection Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do Symptom Cause What To DoTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications Spare Parts List QTY DescriptionTrace Oxygen Analyzer Appendix Drawing ListCalibration Data Calibration ConsiderationsDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix From 288-308K Ambient Temperature K Vapor Pressure of Water mm HgVapor Pressure of Water