5 Maintenance & Troubleshooting | Model 356WA |
|
|
|
|
Symptom | Cause | What To Do |
Cell lacks sensitivity (contin- | Faulty humidity control (contin- | |
ued.) | ||
ued.) | ||
|
To replace humidifier control:
•Shut off all power and remove the AC power cord from the power source.
•Refer to wiring diagram D- 18633 (upper
To replace humidifier heater element:
•Shut off all power and remove the AC power cord from the power source.
•Refer to wiring diagram C- 18593
•Pull out the block assem- bly from the bottom of the humidifier column; pull out the heater element, and unsolder the wires.
•Replace with new heater element (P/N R139.) Solder wires, turn the terminals down into the holder, and replace the holder.
d)Cell compartment not at proper temperature (120 °F).
e)Faulty triac (Q6) in heater control or faulty triac driver (A4).
d)Verify that the power switch is ON.
Check condition of 2A fuse.
e)• Remove the cell compart- ment cover. Refer to Interconnection diagram
•Connect voltmeter (set to high AC) to terminals 1 and 2 of terminal strip TS4 (Schematic diagram
Teledyne Analytical Instruments |