Trace Oxygen Analyzer | Appendix |
|
|
|
|
Appendix
Specifications
Standard Ranges: | Three ranges between |
| |
| able as an option). |
Sampling System: | Wetted parts: 304 welded stainless |
| steel. |
Sensitivity: | 1% of low range. |
Accuracy: | +2% of low range. |
Response and Recovery Time: | 90% in less than 1 minute (for lowest |
| range). |
Operating Temperature Range: | +40 °F to +120 °F (+5 °C to +49 °C). |
Sample Requirement: | Flow: 150 cc/min.* |
| Pressure: 1 to 150 psig. |
| Temperature: +60 °F to +100 °F |
| (+15 °C to +38 °C) |
Power Requirement: | 115 VAC, 50/60 Hz, 150 W (other |
| voltages available.) |
Alarms: | Adjustable single or dual alarm (op- |
| tional) setpoints. |
Recorder Signal Output: | Voltage: |
| Current: |
| (optional.) |
Local Readout: | Digital or analog meter. |
*Specified flow rate required only during calibration. Measuring cell is not sensitive to changes in flow rate.
Teledyne Analytical Instruments |