Teledyne 356WA instruction manual Cell, Electrolyte Replacement, Lead Electrode

Page 24

5 Maintenance & Troubleshooting

Model 356WA

 

 

 

 

5.4 Calibration

The sensitivity of the unit should be checked at two to four week

intervals. Calibration procedures are described in Section 4.7: Calibration.

5.5 Cell

The electrochemical reaction in the analytical process results in the accumulation of lead ions in the electrolyte, making the replacement of the lead electrode or the electrolyte necessary.

5.5.1 Electrolyte Replacement

Replace the electrolyte at least every two months, or even more fre- quently if foreign matter is accumulating in the cell. Remove the cell from is compartment, and drain, clean, rinse, and refill as described in Section 4.2.3: Cell Installation. After the cell is serviced or replaced, calibrate the analyzer as outlined in Section 4.7: Calibration.

5.5.2 Lead Electrode

If electrode is discolored when new, or has obviously deteriorated from use, it may be necessary to clean or replace the electrode. Use the following procedure while referring to Figure 5-1.

If the lead electrode is simply discolored, clean the entire cell according to the following procedure:

a)Heat a quart of Teledyne cleaning solution to slightly below the boiling point, and completely fill the cell cavity with the heated solution.

b)Let the solution stand for approximately five minutes. Drain and dispose of solution in an approved manner.

c)Repeat steps a) and b).

d)Rinse the cell with distilled water and then fill with electrolyte. Let stand for approximately two minutes and then drain and dispose of electrolyte in an approved manner.

e)Refill the cell with electrolyte, immersing the lower edge of the silver screens to about a 3/32 " depth.

If the lead electrode is obviously beyond repair, it must be replaced.

a)Remove the terminal nuts by removing them from the mounting screws.

b)Carefully remove the screw which holds the screen assembly in place. Remove the screen assembly. Use clean tweezers and do

5-2

Teledyne Analytical Instruments

Image 24
Contents Trace Oxygen Analyzer Teledyne Analytical InstrumentsModel 356WA Copyright 1999 Teledyne Analytical InstrumentsTrace Oxygen Analyzer Table of Contents Introduction InstallationMaintenance & Troubleshooting Operational TheoryAppendix Trace Oxygen Analyzer Introduction Method of OperationRequired Equipment Sample ConditioningRecorder /Meter Readout Sensor HumidifierFlow System Flow System Schematic Operational Theory Trace Oxygen Analyzer Installation LocationElectrical Connections Sample Connections Typical System LayoutGND Gas Connections to Back of Analyzer Trace Oxygen Analyzer Operations Filling the ReservoirDetector Cell Cell PackagingTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Throttle Valve Humidity ControlCalibrator Assembly Power Warm-Up and StabilizationCalibration Operations Flowmeter and Humidifier Cell Electrolyte LevelReservoir Electrolyte Replacement CellLead Electrode Cell Assembly Screen Assembly Reservoir and Humidifier ColumnTo Reinstall Removing the Humidifier ColumnLeak Detection Procedure Leak DetectionCell Leak Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do Symptom Cause What To DoTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications Spare Parts List QTY DescriptionTrace Oxygen Analyzer Appendix Drawing ListCalibration Data Calibration ConsiderationsDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix Ambient Temperature K Vapor Pressure of Water mm Hg Vapor Pressure of WaterFrom 288-308K