Teledyne 356WA instruction manual Symptom Cause What To Do

Page 31

 

Trace Oxygen Analyzer

Maintenance & Troubleshooting 5

 

 

 

 

 

 

 

 

 

 

 

 

 

Symptom

 

Cause

 

What To Do

 

Cell lacks sensitivity (contin-

Faulty humidity control (contin-

Turn humidifier control

ued.)

 

ued.)

 

completely counter-

 

 

 

 

 

clockwise.

 

 

 

 

• Turn on main power.

 

 

 

 

While watching the

 

 

 

 

 

voltmeter, turn the

 

 

 

 

 

humidifier control clock-

 

 

 

 

 

wise; the meter reading

 

 

 

 

 

should go from 0–50

 

 

 

 

 

milliamps. If not, replace

 

 

 

 

 

the humidifier heater

 

 

 

 

 

element or the humidifier

 

 

 

 

 

heater control. To deter-

 

 

 

 

 

mine which should be

replaced, see paragraph below.

If there is no current, either the humidifier element or control is defective. To determine which it is:

• Turn the humidifier control knob fully ccw.

• Disconnect the voltmeter from TS1-5 and orange wire. Reconnect the orange wire.

• Change the voltmeter function to 200V AC.

• Place the voltmeter across TS1-5 and termi- nal TS1-4.

• Watching the meter, turn the humidifier control knob cw and watch the voltmeter read 0–110 to 120V AC.

• If not, replace R2 (behind the control knob); see paragraph below.

• If there is voltage, replace

the heater element; see paragraph below.

Teledyne Analytical Instruments

5-9

Image 31
Contents Teledyne Analytical Instruments Trace Oxygen AnalyzerCopyright 1999 Teledyne Analytical Instruments Model 356WAOperational Theory Trace Oxygen Analyzer Table of Contents IntroductionInstallation Maintenance & TroubleshootingAppendix Method of Operation Trace Oxygen Analyzer IntroductionSample Conditioning Required EquipmentRecorder /Meter Readout Humidifier SensorFlow System Flow System Schematic Operational Theory Location Trace Oxygen Analyzer InstallationElectrical Connections Typical System Layout Sample ConnectionsGND Gas Connections to Back of Analyzer Cell Packaging Trace Oxygen Analyzer OperationsFilling the Reservoir Detector CellTypical Model 306WA With 2 Alarm Option Cell Installation Cell Compartment Components Humidity Control Throttle ValveCalibrator Assembly Warm-Up and Stabilization PowerCalibration Operations Cell Electrolyte Level Flowmeter and HumidifierReservoir Cell Electrolyte ReplacementLead Electrode Cell Assembly Reservoir and Humidifier Column Screen AssemblyRemoving the Humidifier Column To ReinstallLeak Detection Leak Detection ProcedureCell Leak Trace Oxygen Analyzer Teledyne Analytical Instruments Symptom Cause What To Do What To Do Symptom CauseTrace Oxygen Analyzer Maintenance & Troubleshooting Trace Oxygen Analyzer Appendix Specifications QTY Description Spare Parts ListDrawing List Trace Oxygen Analyzer AppendixCalibration Considerations Calibration DataDegrees C = 5 F Degrees K = C + Pmm-Hg = 760 2.50 per 100 ft. of altitude Trace Oxygen Analyzer Appendix Vapor Pressure of Water Ambient Temperature K Vapor Pressure of Water mm HgFrom 288-308K