2.4 Calibration

Calibration involves using a span gas to set the span of the instrument. The proper span gas concentration depends on the range that the instrument will be used on. The correct concentration should be 80- 90% of the range used. For instance, if the analyzer is to be used on a range of 0-150 ppm oxygen, then a span gas should be prepared with 120-135 ppm oxygen in nitrogen.

2.4.1 Calibration Procedure for Trace Analysis

To calibrate the Model 3110, the instrument must be fitted with a trace sensor. The instrument must also be purged overnight using a zero gas (a pure gas with no oxygen, typically O2 free N2) before calibrating. For A-2C type cells, match the CO2 content in the span gas to that of the sample gas.

To calibrate the analyzer for trace analysis:

1.Purge the analyzer overnight.

2.Purge the calibration gas sample line, regulator, and control valve. See Section 2.4.3.

3.Set the calibration gas flow rate to 1 SCFH.

4.Attach the vent line followed by the span gas line using the quick disconnect fittings.

5.Navigate to the SPAN VALUE screen (see Section 3.2.13) and set the span value to the known oxygen concentration in ppm of the span gas.

6.Observe the oxygen reading on the screen to determine when the reading has stabilized.

7.Navigate to the SPAN screen and select SPAN: START.

8.When the screen changes and displays SPAN: FINISH, select

SPAN: FINISH to set the span.

9.Allow the span gas to flow for several minutes to verify the proper span setting.

10.Calibration is complete. Remove the span gas line first followed by the vent line.

Teledyne Analytical Instruments

21

Image 21 Contents
Teledyne Analytical Instruments ModelIi Teledyne Analytical Instruments ModelPortable Trace Oxygen Analyzer Specific Model InformationIv Teledyne Analytical Instruments Safety MessagesPortable Trace Oxygen Analyzer Table of Contents Maintenance & Troubleshooting AppendixIndex Teledyne Analytical Instruments Viii Teledyne Analytical Instruments List of FiguresCombustible GAS Usage Introduction IntroductionIntroduction Model Portable Trace Oxygen Analyzer Introduction FeaturesIntroduction Model Cell Warranty Method of AnalysisMicro-Fuel Cell Signal Output Accuracy and ResponseCompact Packaging Portable Trace Oxygen Analyzer Introduction Intentionally left blank Installation Portable Trace Oxygen Analyzer Installation Charging the Batteries Gas Connections Installation ModelPortable Trace Oxygen Analyzer Installation Installing a Trace Sensor Sensor InstallationCalibration Calibration Procedure for Trace AnalysisBleeding the Regulator and Purging the Gas Line Set the Sample Flow rateExternal Signal Front Panel Interface OperationOperation Model Enter KeyPortable Trace Oxygen Analyzer Operation Escape Key3 UP/DOWN Keys Available Menus and Their Sequence Operation and Setup Screens and MenusPower on Screen Home ScreenDate Screen Time Screen EnterAlarm Enable Standard Alarm ScreenHI-LOW Alarm Screen FS Alarm ScreenLtch Alarm Screen Range 2 Screen Range 1 ScreenRange Screen Span Screen Span Value ScreenFilter Screen LOG Intv Screen LOG Reset & Start ScreenLOG Transmit Screen Power Down Screen Portable Trace Oxygen Analyzer Operation Routine Maintenance Maintenance & TroubleshootingReplacing the Battery Maintenance ModelPortable Trace Oxygen Analyzer Maintenance Battery Power Supply ServiceBattery Charger Port on the Model Cell ReplacementCell Warranty Temperature Compensation Effects of CO2 on B-2 Cell Life Effect of CO2 on B-2 Cell LifeProcedure Leak TestingIntentionally left blank Specifications AppendixAppendix Model Portable Trace Oxygen Analyzer Appendix Spare Parts ListIndex Portable Trace Oxygen Analyzer Index Power Down screen, 39 power offAppendix Model