3.2.13 SPAN VALUE Screen

The SPAN VALUE screen displays the oxygen concentration of the span gas used for calibration. This is not a measured value; it is the known span gas concentration that is input to the analyzer by the operator.

To change the span gas concentration:

1.Use the UP/DOWN keys to navigate to the SPAN VALUE screen. Then press ENTER to enter the setup screen.

2.Use the UP /DOWN keys to change the span value. Press ENTER to accept the displayed value and to return to the SPAN VALUE screen.

3.2.14SPAN Screen

This screen is used to perform a span calibration on the Model 3110. The appropriate span value must have already been input to the instrument. See Section 3.2.13 for entering a span value into the analyzer.

CAUTION: THE CORRECT SPAN VALUE MUST BE ENTERED INTO THE INSTRUMENT THAT CORRESPONDS TO THE ACTUAL SPAN GAS USED. FAILURE TO ENTER THE PROPER VALUE WILL RESULT IN ERRONEOUS READINGS.

Note: The filter should be set to a low value to avoid errors during span setting. Allow the oxygen reading to become stable prior to setting the span.

To perform a span calibration:

34 Teledyne Analytical Instruments

Image 34 Contents
Model Teledyne Analytical InstrumentsModel Ii Teledyne Analytical InstrumentsSpecific Model Information Portable Trace Oxygen AnalyzerSafety Messages Iv Teledyne Analytical InstrumentsPortable Trace Oxygen Analyzer Table of Contents Appendix Maintenance & TroubleshootingIndex Teledyne Analytical Instruments List of Figures Viii Teledyne Analytical InstrumentsCombustible GAS Usage Introduction IntroductionIntroduction Model Features Portable Trace Oxygen Analyzer IntroductionIntroduction Model Method of Analysis Cell WarrantyMicro-Fuel Cell Accuracy and Response Signal OutputCompact Packaging Portable Trace Oxygen Analyzer Introduction Intentionally left blank Portable Trace Oxygen Analyzer Installation InstallationCharging the Batteries Installation Model Gas ConnectionsPortable Trace Oxygen Analyzer Installation Sensor Installation Installing a Trace SensorCalibration Procedure for Trace Analysis CalibrationSet the Sample Flow rate Bleeding the Regulator and Purging the Gas LineExternal Signal Operation Model OperationFront Panel Interface Enter KeyEscape Key Portable Trace Oxygen Analyzer Operation3 UP/DOWN Keys Operation and Setup Screens and Menus Available Menus and Their SequenceHome Screen Power on ScreenDate Screen Enter Time ScreenStandard Alarm Screen Alarm Enable FS Alarm Screen HI-LOW Alarm ScreenLtch Alarm Screen Range 1 Screen Range 2 ScreenRange Screen Span Value Screen Span ScreenFilter Screen LOG Reset & Start Screen LOG Intv ScreenLOG Transmit Screen Power Down Screen Portable Trace Oxygen Analyzer Operation Replacing the Battery Maintenance & TroubleshootingRoutine Maintenance Maintenance ModelBattery Power Supply Service Portable Trace Oxygen Analyzer MaintenanceCell Replacement Battery Charger Port on the ModelCell Warranty Temperature Compensation Effect of CO2 on B-2 Cell Life Effects of CO2 on B-2 Cell LifeLeak Testing ProcedureIntentionally left blank Appendix SpecificationsAppendix Model Spare Parts List Portable Trace Oxygen Analyzer AppendixIndex Power Down screen, 39 power off Portable Trace Oxygen Analyzer IndexAppendix Model