Operation | Model 3110 |
|
|
3.2.13 SPAN VALUE Screen
The SPAN VALUE screen displays the oxygen concentration of the span gas used for calibration. This is not a measured value; it is the known span gas concentration that is input to the analyzer by the operator.
To change the span gas concentration:
1.Use the UP/DOWN keys to navigate to the SPAN VALUE screen. Then press ENTER to enter the setup screen.
2.Use the UP /DOWN keys to change the span value. Press ENTER to accept the displayed value and to return to the SPAN VALUE screen.
3.2.14SPAN Screen
This screen is used to perform a span calibration on the Model 3110. The appropriate span value must have already been input to the instrument. See Section 3.2.13 for entering a span value into the analyzer.
CAUTION: THE CORRECT SPAN VALUE MUST BE ENTERED INTO THE INSTRUMENT THAT CORRESPONDS TO THE ACTUAL SPAN GAS USED. FAILURE TO ENTER THE PROPER VALUE WILL RESULT IN ERRONEOUS READINGS.
Note: The filter should be set to a low value to avoid errors during span setting. Allow the oxygen reading to become stable prior to setting the span.
To perform a span calibration: