Teledyne 3110 operating instructions LOG Transmit Screen

Page 37

Portable Trace Oxygen Analyzer

Operation

 

 

RESET menu again. The data set can be downloaded using the LOG TRANSMIT screen (see Section 3.2.17).

To start the data logger:

1.Use the UP/DOWN keys to navigate to the LOG RESET & START screen and press ENTER.

2.When the screen begins to blink press ENTER again to start the data logger. Once the data logger has started, the screen changes RESET&START to STOP. During logging a blinking “L” appears in the upper left corner of the screen.

To stop the data logger:

1.Press ENTER to stop the data logger and return to the LOG RESET & START screen.

3.2.18LOG TRANSMIT Screen

Data can be downloaded to a computer using the optional RS-232 port and cable attached to a PC. The RS-232 port accepts a standard data cable with a DB-9 connector. The computer must be able to accept data from a RS-232 source with the following characteristics:

9600 baud

8 bit data no parity

Teledyne Analytical Instruments

37

Image 37 Contents
Teledyne Analytical Instruments ModelIi Teledyne Analytical Instruments ModelPortable Trace Oxygen Analyzer Specific Model InformationIv Teledyne Analytical Instruments Safety MessagesPortable Trace Oxygen Analyzer Table of Contents Appendix Maintenance & TroubleshootingIndex Teledyne Analytical Instruments Viii Teledyne Analytical Instruments List of FiguresCombustible GAS Usage Introduction IntroductionIntroduction Model Portable Trace Oxygen Analyzer Introduction FeaturesIntroduction Model Method of Analysis Cell WarrantyMicro-Fuel Cell Accuracy and Response Signal OutputCompact Packaging Portable Trace Oxygen Analyzer Introduction Intentionally left blank Portable Trace Oxygen Analyzer Installation InstallationCharging the Batteries Gas Connections Installation ModelPortable Trace Oxygen Analyzer Installation Installing a Trace Sensor Sensor InstallationCalibration Calibration Procedure for Trace AnalysisBleeding the Regulator and Purging the Gas Line Set the Sample Flow rateExternal Signal Front Panel Interface OperationOperation Model Enter KeyEscape Key Portable Trace Oxygen Analyzer Operation3 UP/DOWN Keys Available Menus and Their Sequence Operation and Setup Screens and MenusHome Screen Power on ScreenDate Screen Time Screen EnterAlarm Enable Standard Alarm ScreenHI-LOW Alarm Screen FS Alarm ScreenLtch Alarm Screen Range 2 Screen Range 1 Screen Range Screen Span Screen Span Value ScreenFilter Screen LOG Intv Screen LOG Reset & Start ScreenLOG Transmit Screen Power Down Screen Portable Trace Oxygen Analyzer Operation Routine Maintenance Maintenance & TroubleshootingReplacing the Battery Maintenance ModelPortable Trace Oxygen Analyzer Maintenance Battery Power Supply ServiceBattery Charger Port on the Model Cell ReplacementCell Warranty Temperature Compensation Effects of CO2 on B-2 Cell Life Effect of CO2 on B-2 Cell LifeProcedure Leak TestingIntentionally left blank Appendix SpecificationsAppendix Model Portable Trace Oxygen Analyzer Appendix Spare Parts ListIndex Portable Trace Oxygen Analyzer Index Power Down screen, 39 power offAppendix Model