Teledyne 3110 operating instructions Enter, Time Screen

Page 28

Operation

Model 3110

 

 

2.Use the UP/DOWN keys to alter the month. Then press ENTER. The cursor will move over to the next editable field.

3.Use the UP/DOWN keys to alter the day. Then press ENTER. The cursor will move over to the next editable field.

4.Use the UP/DOWN keys to alter the year. Then press

ENTER.

5.Press ENTER again to save the current date and automatically return to the DATE screen.

At any time you can press the ESC key to abort the entry and return to the DATE screen.

3.2.4 TIME Screen

Use the UP/DOWN keys to navigate to the TIME screen. The second line of the this screen displays the current time in 24 hour military format. This information is used by the data logger for time stamping data records. The currently set time is displayed on the second line of the display.

To change the currently set time:

1.Press ENTER to enter the time setup function. Note the navigation arrows that appear on the left and right sides of the display change from UP/DOWN to pointing

LEFT/RIGHT.

2.Use the UP/DOWN keys to alter the hour field. Then press ENTER. The cursor will move over to the next editable field.

3.Use the UP/DOWN keys to alter the minute field. Then press ENTER. The cursor will move over to the next editable field.

4.Use the UP/DOWN keys to alter the seconds field. Then press ENTER.

5.Press ENTER again to save the current time and automatically return to the TIME screen.

28 Teledyne Analytical Instruments

Image 28 Contents
Model Teledyne Analytical InstrumentsModel Ii Teledyne Analytical InstrumentsSpecific Model Information Portable Trace Oxygen AnalyzerSafety Messages Iv Teledyne Analytical InstrumentsPortable Trace Oxygen Analyzer Table of Contents Appendix Maintenance & TroubleshootingIndex Teledyne Analytical Instruments List of Figures Viii Teledyne Analytical InstrumentsCombustible GAS Usage Introduction IntroductionIntroduction Model Features Portable Trace Oxygen Analyzer IntroductionIntroduction Model Method of Analysis Cell WarrantyMicro-Fuel Cell Accuracy and Response Signal OutputCompact Packaging Portable Trace Oxygen Analyzer Introduction Intentionally left blank Portable Trace Oxygen Analyzer Installation InstallationCharging the Batteries Installation Model Gas ConnectionsPortable Trace Oxygen Analyzer Installation Sensor Installation Installing a Trace SensorCalibration Procedure for Trace Analysis CalibrationSet the Sample Flow rate Bleeding the Regulator and Purging the Gas LineExternal Signal Operation Front Panel Interface Operation Model Enter KeyEscape Key Portable Trace Oxygen Analyzer Operation3 UP/DOWN Keys Operation and Setup Screens and Menus Available Menus and Their SequenceHome Screen Power on ScreenDate Screen Enter Time ScreenStandard Alarm Screen Alarm EnableFS Alarm Screen HI-LOW Alarm ScreenLtch Alarm Screen Range 1 Screen Range 2 ScreenRange Screen Span Value Screen Span ScreenFilter Screen LOG Reset & Start Screen LOG Intv ScreenLOG Transmit Screen Power Down Screen Portable Trace Oxygen Analyzer Operation Maintenance & Troubleshooting Routine MaintenanceReplacing the Battery Maintenance ModelBattery Power Supply Service Portable Trace Oxygen Analyzer MaintenanceCell Replacement Battery Charger Port on the ModelCell Warranty Temperature Compensation Effect of CO2 on B-2 Cell Life Effects of CO2 on B-2 Cell LifeLeak Testing ProcedureIntentionally left blank Appendix SpecificationsAppendix Model Spare Parts List Portable Trace Oxygen Analyzer AppendixIndex Power Down screen, 39 power off Portable Trace Oxygen Analyzer IndexAppendix Model