Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

3.3Electrical Connections

Figure 3-1 shows the Model 3300TB rear panel of the control unit. For detailed pinouts, see the wiring/interconnection drawings in the Draw- ings section at the rear of this manual. To gain access to the terminal blocks, the analyzer door screws must be lossened, and the door opened.

Figure 3-1 Electrical Connectors for AC Control Unit

Primary Input Power: The power strip supplied inside the analyzer. Connect ground to terminal 1, neutral to terminal 3, and hot to terminal 2. Make sure female plug end is inserted in the control unit power receptacle.

The universal power supply allows direct connection to any 100-240 VAC, 50/60Hz power source. The fuse block, to the right of the power cord receptacle, accepts two 5x20mm 0.5 A, 250V, IEC time-lag (T) fuse. (See Fuse Replacement in chapter 5, Maintenance.)

The Power switch is located on the right-hand end of the power source input receptacle assembly.

AC Power Terminal Strip

Teledyne Analytical Instruments

3-3

Image 19
Contents Model 3300TB Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceProtective Earth Combustible GAS Usage Warning Introduction Overview Main Features of the AnalyzerTrace Oxygen Analyzer Introduction Front Panel Description Set AlarmIntroduction Model 3300TB Rear Panel Rear Panel DescriptionRS-232 Port Analog OutputsPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3300TBEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsElectronics Signal ProcessingGeneral Operational Theory Model 3300TB Trace Oxygen Analyzer Installation InstallationUnpacking the Analyzer Location and Mounting Installation Model 3300TBControl Unit Installation Installing the Micro-Fuel Cell / Cell Block OrientationElectrical Connections Electrical Connectors for AC Control UnitContact ID for Failsafe Relay Operation Sensor Fail Alarm #1Alarm #2 Gas ConnectionsPiping Diagram for Vacuum Service Option Installation Model 3300TB Vacuum Service OptionInstallation Checklist Installation Model 3300TB Trace Oxygen Analyzer Operation Operation IntroductionSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300TB LO Range Setting the Alarm SetpointsSet Alarm Trace Oxygen Analyzer Operation HI RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation Model 3300 TB Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance Remove Power to Unit before replacing the fuseOrdering and Handling of Spare Sensors Sensor Installation or ReplacementMaintenance Model 3300TB When to Replace a SensorExploded View of MFC and Cell Block Removing the Micro-Fuel CellDisconnect the Power Source from the Control Unit Installing a Micro-Fuel CellCell Warranty Conditions Trace Oxygen Analyzer Maintenance Maintenance Model 3300TB Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300TB Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300TB Material Safety Data Sheet Effects of overexposure Ingestion Signs/symptoms of exposurePrimary route of entry Exposure limits Osha PELDisposal Protective measures