Teledyne 3300TB operating instructions Replacing the Fuse, Trace Oxygen Analyzer Maintenance

Page 31

Trace Oxygen Analyzer

Maintenance 5

 

 

 

 

 

 

 

 

 

Maintenance

Aside from normal cleaning and checking for leaks at the gas connections, the Model 3300TB should not require any maintenance beyond replacement of expended Micro-Fuel Cells, and perhaps a blown fuse. Routine maintenance includes occasional recalibration, as described in chapter 4, Operation.

5.1Replacing the Fuse

Remove Power to Unit before replacing the fuse.

When a fuse blows, check first to determine the cause, then replace the fuse using the following procedure:

1.Disconnect the AC power and place the power switch located on the rear panel in the O position. Remove the power cord from the receptacle.

2.The fuse receptacle is located in the power cord receptacle assembly in the upper left-hand corner of the rear panel. See Figure 5-1.

FUSE

Figure 5-1: AC Fuse Replacement

3.Insert a small flat-blade screwdriver into the slot in the receptacle wall nearest the fuse and gently pry open the fuse receptacle. The

Teledyne Analytical Instruments

5-1

Image 31
Contents Model 3300TB Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceProtective Earth Combustible GAS Usage Warning Introduction Overview Main Features of the AnalyzerTrace Oxygen Analyzer Introduction Front Panel Description Set AlarmIntroduction Model 3300TB Rear Panel Rear Panel DescriptionRS-232 Port Analog OutputsPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3300TBEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsElectronics Signal ProcessingGeneral Operational Theory Model 3300TB Trace Oxygen Analyzer Installation InstallationUnpacking the Analyzer Location and Mounting Installation Model 3300TBControl Unit Installation Installing the Micro-Fuel Cell / Cell Block OrientationElectrical Connections Electrical Connectors for AC Control UnitContact ID for Failsafe Relay Operation Sensor Fail Alarm #1Alarm #2 Gas ConnectionsPiping Diagram for Vacuum Service Option Installation Model 3300TB Vacuum Service OptionInstallation Checklist Installation Model 3300TB Trace Oxygen Analyzer Operation Operation IntroductionSetting the Analysis Ranges Using the Function and Data Entry ButtonsOperation Model 3300TB LO Range Setting the Alarm SetpointsSet Alarm Trace Oxygen Analyzer Operation HI RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingSupplementary Information Displaying Percent & PPM on the LED DisplayOperation Model 3300 TB Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance Remove Power to Unit before replacing the fuseOrdering and Handling of Spare Sensors Sensor Installation or ReplacementMaintenance Model 3300TB When to Replace a SensorExploded View of MFC and Cell Block Removing the Micro-Fuel CellDisconnect the Power Source from the Control Unit Installing a Micro-Fuel CellCell Warranty Conditions Trace Oxygen Analyzer Maintenance Maintenance Model 3300TB Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300TB Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300TB Material Safety Data Sheet Effects of overexposure Ingestion Signs/symptoms of exposurePrimary route of entry Exposure limits Osha PELDisposal Protective measures