Teledyne 3300TB operating instructions Installation Checklist

Page 23

Trace Oxygen Analyzer

Installation 3

 

 

 

 

 

 

 

 

 

3.5Installation Checklist

Before connecting the instrument to the power source and turning it on, make sure you have:

Correctly installed the Sample and Exhaust gas lines

Checked for leaks

Set the sample pressure to 5–10 psig (34.5 - 68.9 kPa), nominal (for non-vacuum service units)

Set the flow

Once the above checks have been made, you can connect to the power source. The instrument is now ready for operation.

Teledyne Analytical Instruments

3-7

Image 23
Contents Model 3300TB Copyright 1999 Teledyne Analytical Instruments Contents Appendix MaintenanceProtective Earth Combustible GAS Usage Warning Trace Oxygen Analyzer Introduction Main Features of the AnalyzerIntroduction Overview Introduction Model 3300TB Set AlarmFront Panel Description Rear Panel Rear Panel DescriptionRS-232 Port Analog OutputsPrinciples of Operation Operational Theory IntroductionMicro-Fuel Cell Sensor Trace Oxygen Analyzer Operational TheoryAnatomy of a Micro-Fuel Cell Operational Theory Model 3300TBEffect of Pressure Electrochemical ReactionsCharacteristic Input/Output Curve for a Micro-Fuel Cell Calibration CharacteristicsGeneral Signal ProcessingElectronics Operational Theory Model 3300TB Unpacking the Analyzer InstallationTrace Oxygen Analyzer Installation Location and Mounting Installation Model 3300TBControl Unit Installation Installing the Micro-Fuel Cell / Cell Block OrientationElectrical Connections Electrical Connectors for AC Control UnitContact ID for Failsafe Relay Operation Sensor Fail Alarm #1Alarm #2 Gas ConnectionsPiping Diagram for Vacuum Service Option Installation Model 3300TB Vacuum Service OptionInstallation Checklist Installation Model 3300TB Trace Oxygen Analyzer Operation Operation IntroductionOperation Model 3300TB Using the Function and Data Entry ButtonsSetting the Analysis Ranges LO Range Setting the Alarm SetpointsSet Alarm Trace Oxygen Analyzer Operation HI RangeSensor Fail Alarm Calibration Selecting a Fixed Range or AutorangingOperation Model 3300 TB Displaying Percent & PPM on the LED DisplaySupplementary Information Replacing the Fuse MaintenanceTrace Oxygen Analyzer Maintenance Remove Power to Unit before replacing the fuseOrdering and Handling of Spare Sensors Sensor Installation or ReplacementMaintenance Model 3300TB When to Replace a SensorExploded View of MFC and Cell Block Removing the Micro-Fuel CellCell Warranty Conditions Installing a Micro-Fuel CellDisconnect the Power Source from the Control Unit Trace Oxygen Analyzer Maintenance Maintenance Model 3300TB Trace Oxygen Analyzer Appendix Appendix SpecificationsAppendix Model 3300TB Spare Parts ListMiscellaneous Reference DrawingAppendix Model 3300TB Material Safety Data Sheet Effects of overexposure Ingestion Signs/symptoms of exposurePrimary route of entry Exposure limits Osha PELDisposal Protective measures