2Using the Handheld Infrared Thermometer

2.3.1Adjusting Emissivity

°F

Refer to Appendices B and C for information on emissivity.

1.Determine the emissivity of the target.

2.Aim at the target and pull the trigger.

3.If necessary, press the key to increment the target

emissivity or press the key to decrement the target emissivity.

NOTE

The Emissivity Display Mode (E) appears every time the trigger is pulled regardless of how the Display Mode was previously set.

NOTE

The emissivity setting does not change when the thermometer is turned off. However, when the batteries are replaced, the emissivity is reset to 0.95, the default value.

2.3.2 Using the LOCK Function

This function electronically locks the trigger mechanism:

1.Pull the trigger and press the key to lock the trigger. The

icon will appear on the display.

2.Release the trigger. This allows the thermometer to operate continuously whether or not the trigger is pulled.

NOTE

To unlock the trigger function, press the key again,

and the icon is no longer displayed.

2-12

Image 27
Contents User’s Guide Servicing North America For immediate technical or application assistanceServicing Europe Unpacking Instructions Model No Description Page Unpacking Instructions Chapter General Description Contents General Description Introduction Features OS531 OS532 OS533 OS534 OS530HR OS530L-CF Features OS523 OS524 Front of the Thermometer Parts of the ThermometerDisplay Details Key OS530 Series Handheld Infrared Thermometer Rear View Rear of the ThermometerGeneral Description Ac Power Operation Battery OperationOS530/OS520 Series with Built-in Laser Sighting Operating the ThermometerField of View OS531, OS532, OS530L Field of View OS534, OS523-1 Field of View OS523-2 10 Field of View OS524 Trigger Measurement Techniques11 Recorder Hookup If the trigger is pulled two times in rapid sequence Real Time Mode Active OperationRecall Mode 13. Visual Function Flow Chart Using the Lock Function Adjusting EmissivityCalculating Temperature Values Changing the Temperature from F to C or vice versaTurning on the Display Backlighting Thermocouple Input OS532, OS533, OS534 If you are not in High Alarm Display Mode HAL when Using the Alarm FunctionsTo set the low alarm value OS533, OS534, OS523, OS524 To disable this mode, press the key again Icon disappears One Stop Bit Key to decrement the printing interval. System Requirements Irtempsoft Connecting the IR-System to the PC Using the Handheld Infrared Thermometer Storing the Temperature Data on Command OS534, OS523, OS524 Erasing the Temperature Data from Memory Mode Recall Mode Passive OperationFunctional Flow Chart Recall Mode Reviewing the Last Parameters AMB temperature MEM locationPRN F Using the Handheld Infrared Thermometer Laser Sighting Rear View DescriptionLaser Dot Operating the Laser SightingLines of Sight of the Laser Sighting and Thermometer Sighting Scope Installing and Operating the Sighting ScopeInstalling the Sighting Scope Maintenance Replacing the Batteries Safety WarningCleaning the Lens Calibrating the ThermometerServicing the Laser Sighting Problem Solution Troubleshooting GuideProblem If you see an error code, either Target are offset by design Specifications Specifications are for all models except where notedOS533, OS534, OS523, OS524 No Laser Sighting 17 hours continuous Laser Sighting Glossary of Key Strokes Glossary of Key Strokes Appendix How Infrared Thermometry Works a Thermal RadiationBlackbody Calculating Temperature Optics Field of View Appendix Emissivity Values MetalsMaterial Emissivity ε NonmetalsAppendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER PH/CONDUCTIVITY Shop online at omega.com