2Using the Handheld Infrared Thermometer

ATC°F

8.Press and hold the key until the Emissivity Display Mode (E) appears.

9.Change the emissivity to the proper value for the target being measured (refer to Section 2.3.1).

10.Aim at the target. The target temperature and emissivity are displayed on the LCD.

11.After all data is taken, press the key to release this mode

NOTE

To disable the Ambient Target Temperature

Compensation at a later time, you must press the

key to get into the Ambient Target Temperature Display Mode. Then press the key to disable it.

NOTE

The target ambient temperature does not change when the thermometer is turned off. However, when the batteries are replaced, it is reset to the default value of 75°F.

2.3.9Sending Temperature Data to a Serial Printer (OS533, OS534, OS523, OS524)

PRN °F

The thermometer can transmit temperature data to a Serial Printer via the RS-232 phone jack and the RS-232 cable.

1.Turn on the Serial Printer and set the following parameters:

Speed:

4800 BPS

One Stop Bit

Data:

8 Bits

No Parity

2.Connect the RS-232 cable between the thermometer and the printer as shown in Figure 2-14.

2-18

Image 33
Contents User’s Guide Servicing North America For immediate technical or application assistanceServicing Europe Unpacking Instructions Model No Description Page Unpacking Instructions Chapter General Description Contents General Description Introduction Features OS531 OS532 OS533 OS534 OS530HR OS530L-CF Features OS523 OS524 Front of the Thermometer Parts of the ThermometerDisplay Details Key OS530 Series Handheld Infrared Thermometer Rear View Rear of the ThermometerGeneral Description Ac Power Operation Battery OperationOS530/OS520 Series with Built-in Laser Sighting Operating the ThermometerField of View OS531, OS532, OS530L Field of View OS534, OS523-1 Field of View OS523-2 10 Field of View OS524 Trigger Measurement Techniques11 Recorder Hookup If the trigger is pulled two times in rapid sequence Real Time Mode Active OperationRecall Mode 13. Visual Function Flow Chart Using the Lock Function Adjusting EmissivityCalculating Temperature Values Changing the Temperature from F to C or vice versaTurning on the Display Backlighting Thermocouple Input OS532, OS533, OS534 If you are not in High Alarm Display Mode HAL when Using the Alarm FunctionsTo set the low alarm value OS533, OS534, OS523, OS524 To disable this mode, press the key again Icon disappears One Stop Bit Key to decrement the printing interval. System Requirements Irtempsoft Connecting the IR-System to the PC Using the Handheld Infrared Thermometer Storing the Temperature Data on Command OS534, OS523, OS524 Erasing the Temperature Data from Memory Mode Recall Mode Passive OperationFunctional Flow Chart Recall Mode Reviewing the Last Parameters AMB temperature MEM locationPRN F Using the Handheld Infrared Thermometer Laser Sighting Rear View DescriptionLaser Dot Operating the Laser SightingLines of Sight of the Laser Sighting and Thermometer Sighting Scope Installing and Operating the Sighting ScopeInstalling the Sighting Scope Maintenance Replacing the Batteries Safety WarningCleaning the Lens Calibrating the ThermometerServicing the Laser Sighting Problem Solution Troubleshooting GuideProblem If you see an error code, either Target are offset by design Specifications Specifications are for all models except where notedOS533, OS534, OS523, OS524 No Laser Sighting 17 hours continuous Laser Sighting Glossary of Key Strokes Glossary of Key Strokes Appendix How Infrared Thermometry Works a Thermal RadiationBlackbody Calculating Temperature Optics Field of View Appendix Emissivity Values MetalsMaterial Emissivity ε NonmetalsAppendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER PH/CONDUCTIVITY Shop online at omega.com