Omega Engineering OS532, OS530HR, OS531 To disable this mode, press the key again Icon disappears

Page 32

Using the Handheld Infrared Thermometer 2

2.3.8Using Ambient Target Temperature Compensation (OS533, OS534, OS523, OS524)

ATC°F

Use the Ambient Target Temperature Compensation

(AMB) Display Mode when high accuracy readings under both of these conditions are required:

The target has a low emissivity.

The ambient temperature around the target is much higher than the ambient temperature around the infrared thermometer.

To set and activate the Ambient Target Temperature Compensation Mode:

1.Pull the trigger and press the key to lock the trigger. Set the emissivity to 1.0 (refer to Section 2.3.1).

2.Press and hold the key until the Average Display Mode (AVG) appears.

3.Slowly move the thermometer so that the line of sight sweeps across the area surrounding the target. The thermometer measures the temperature at each point on the surrounding area.

4.Read the average temperature value from the upper display and record it here __________.

5.Press and hold the key until the Ambient Temperature Display Mode (AMB) appears.

6.Set the AMB temperature found in Step 4 by pressing the

key or the key.

7.Press the key to enable the ambient target

temperature compensation. The icon appears on the display.

NOTE

To disable this mode, press the key again.

The icon disappears.

2-17

Image 32
Contents User’s Guide Servicing Europe Servicing North AmericaFor immediate technical or application assistance Unpacking Instructions Model No Description Page Unpacking Instructions Chapter General Description Contents General Description Introduction Features OS531 OS532 OS533 OS534 OS530HR OS530L-CF Features OS523 OS524 Parts of the Thermometer Front of the ThermometerDisplay Details Key Rear of the Thermometer OS530 Series Handheld Infrared Thermometer Rear ViewGeneral Description Battery Operation Ac Power OperationOperating the Thermometer OS530/OS520 Series with Built-in Laser SightingField of View OS531, OS532, OS530L Field of View OS534, OS523-1 Field of View OS523-2 10 Field of View OS524 Measurement Techniques Trigger11 Recorder Hookup Real Time Mode Active Operation If the trigger is pulled two times in rapid sequenceRecall Mode 13. Visual Function Flow Chart Adjusting Emissivity Using the Lock FunctionTurning on the Display Backlighting Calculating Temperature ValuesChanging the Temperature from F to C or vice versa Thermocouple Input OS532, OS533, OS534 Using the Alarm Functions If you are not in High Alarm Display Mode HAL whenTo set the low alarm value OS533, OS534, OS523, OS524 To disable this mode, press the key again Icon disappears One Stop Bit Key to decrement the printing interval. System Requirements Irtempsoft Connecting the IR-System to the PC Using the Handheld Infrared Thermometer Storing the Temperature Data on Command OS534, OS523, OS524 Erasing the Temperature Data from Memory Recall Mode Passive Operation ModeFunctional Flow Chart Recall Mode AMB temperature MEM location Reviewing the Last ParametersPRN F Using the Handheld Infrared Thermometer Laser Sighting Description Rear ViewOperating the Laser Sighting Laser DotLines of Sight of the Laser Sighting and Thermometer Installing and Operating the Sighting Scope Sighting ScopeInstalling the Sighting Scope Safety Warning Maintenance Replacing the BatteriesServicing the Laser Sighting Cleaning the LensCalibrating the Thermometer Troubleshooting Guide Problem SolutionProblem If you see an error code, either Target are offset by design Specifications are for all models except where noted SpecificationsOS533, OS534, OS523, OS524 No Laser Sighting 17 hours continuous Laser Sighting Glossary of Key Strokes Glossary of Key Strokes Thermal Radiation Appendix How Infrared Thermometry Works aBlackbody Calculating Temperature Optics Field of View Metals Appendix Emissivity ValuesNonmetals Material Emissivity εAppendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER Shop online at omega.com PH/CONDUCTIVITY