2Using the Handheld Infrared Thermometer
2.3.11Storing the Temperature Data on Command (OS534, OS523, OS524)
°F
The thermometer can store up to 100 temperature data points on command. Each set of temperature data is broken down into the temperature value, emissivity, and high alarm setpoint for that temperature. This data is stored in
1.Aim at the target and pull the trigger and press the
key to lock the trigger. The icon will appear on the display.
2.If necessary, press the key to increment the target
emissivity or press the key to decrement the target emissivity.
3.Press and hold the key until the Memory Display Mode (MEM) appears.
4.Press the key to increment the memory location or
press the key to decrement the memory location. The memory location can be from 001 to 100.
5.Press the key to store the target temperature at the memory location indicated. You will hear a beep to
verify that the data is stored.
NOTE
You can write over any previously stored memory locations.
6.After all data is taken, press and hold the key until the Emissivity Display Mode (E) appears.
7.Press the key to unlock the trigger.