Using the Handheld Infrared Thermometer 2

2.3.7 Using the Alarm Functions

HAL

°F

The thermometer provides audible and visible alarm indications.

To set the high alarm value:

1.Pull the trigger. Then press and hold the key until the High Alarm Display Mode (HAL) appears.

2.Press the key to increment the high alarm value. Press

the key to decrement the high alarm value.

3.Press the key to enable the high alarm function. The

icon appears.

If the temperature exceeds the high alarm setpoint,

you will hear a beep and the icon on the display flashes.

4.To disable the high alarm, press the key again,

and the icon disappears.

NOTE

If you are not in High Alarm Display Mode (HAL) when

the high alarm goes off, you must press the key to get into the High Alarm Display Mode. Then press the

key to disable the high alarm.

NOTE

The high alarm setpoint does not change when the thermometer is turned off. However, when the batteries are replaced, it is reset to the default value as follows:

OS530HR:250°F

OS531:750°F

OS530L, OS532, OS533: 1000°F

OS534:1600°F

OS523:2500°F

OS524:4500°F

2-15

Image 30
Contents User’s Guide Servicing North America For immediate technical or application assistanceServicing Europe Unpacking Instructions Model No Description Page Unpacking Instructions Chapter General Description Contents General Description Introduction Features OS531 OS532 OS533 OS534 OS530HR OS530L-CF Features OS523 OS524 Parts of the Thermometer Front of the ThermometerDisplay Details Key Rear of the Thermometer OS530 Series Handheld Infrared Thermometer Rear ViewGeneral Description Battery Operation Ac Power OperationOperating the Thermometer OS530/OS520 Series with Built-in Laser SightingField of View OS531, OS532, OS530L Field of View OS534, OS523-1 Field of View OS523-2 10 Field of View OS524 Measurement Techniques Trigger11 Recorder Hookup Real Time Mode Active Operation If the trigger is pulled two times in rapid sequenceRecall Mode 13. Visual Function Flow Chart Adjusting Emissivity Using the Lock FunctionCalculating Temperature Values Changing the Temperature from F to C or vice versaTurning on the Display Backlighting Thermocouple Input OS532, OS533, OS534 Using the Alarm Functions If you are not in High Alarm Display Mode HAL whenTo set the low alarm value OS533, OS534, OS523, OS524 To disable this mode, press the key again Icon disappears One Stop Bit Key to decrement the printing interval. System Requirements Irtempsoft Connecting the IR-System to the PC Using the Handheld Infrared Thermometer Storing the Temperature Data on Command OS534, OS523, OS524 Erasing the Temperature Data from Memory Recall Mode Passive Operation ModeFunctional Flow Chart Recall Mode AMB temperature MEM location Reviewing the Last ParametersPRN F Using the Handheld Infrared Thermometer Laser Sighting Description Rear ViewOperating the Laser Sighting Laser DotLines of Sight of the Laser Sighting and Thermometer Installing and Operating the Sighting Scope Sighting ScopeInstalling the Sighting Scope Safety Warning Maintenance Replacing the BatteriesCleaning the Lens Calibrating the ThermometerServicing the Laser Sighting Troubleshooting Guide Problem SolutionProblem If you see an error code, either Target are offset by design Specifications are for all models except where noted SpecificationsOS533, OS534, OS523, OS524 No Laser Sighting 17 hours continuous Laser Sighting Glossary of Key Strokes Glossary of Key Strokes Thermal Radiation Appendix How Infrared Thermometry Works aBlackbody Calculating Temperature Optics Field of View Metals Appendix Emissivity ValuesNonmetals Material Emissivity εAppendix Determining an Unknown Emissivity Figure C-1. Determining Emissivity Figure C-2. Determining Emissivity with a Drilled Hole Appendix Determining an Unknown Emissivity Index Field of View Diagrams Recall WARRANTY/DISCLAIMER Shop online at omega.com PH/CONDUCTIVITY